56
(7)Saliva Ejector SE-90QT
1)Operation
The saliva ejector begins to operate when it is taken off its holder on
the assistant’s side and stops when it is returned to its holder.
The ejector will not operate if it is pulled out while
the auto cup filler is operating.
2)Nozzle Connection
Slide the SE nozzle straight into the rubber end of the ejector until it is
held firmly in place.
3)Body Detachment
The body can be taken off for autoclaving.
Click the metal part and pull in straight off.
To reattach it, pull it straight on until an audible click is heard..
(8)Ultrasonic Scaler
*
1)Operation
Step on the foot pedal to operate the scaler. It vibrates at a steady
rate.
When the spray is turned on, it will be emitted when the foot pedal is
depressed.
There are 10 power levels for each power range. The upper limit for
each range can be set. (See 2) below.)
The vibration level is shown in the display. (Refer to level estimate.)
If the scaler is set for continuo vibration, stepping on the pedal will
start it, and stepping on the pedal again will stop it.(See 2) below.)
ADVICE
Ejector
Nozzle
Ejector
Ejector
Body
[5]Operation 6.(7)(8)
Power Level Estimates
Setting Power Range
Weak
(
periodontal
):
Medium
(
Endodontic
):
Strong
(
scaling
):
Weakest 0 to
medium 10
Weak 0 to
medium 10
medium 0 to
strongest
*: Option