60
B
5)Flushing Device
*
This is used to flush out all the water lines inside the unit.
Do this once a day.
Switch On the main switch and follow the direction in the display.
a)Set the flushing bowl up on the basin. Line up and fit the gauge on
top of the bowl to the end of the auto filler for correct positioning.
There is a magnet in the bottom of the bowl, and
a sensor detects whether the bowl is correctly
positioned.
When the bowl is in the correct position, the auto
filler stops running and “Flushing” appears in the
display.
b)Connect the main tubes to the flushing bowl.
Do not fail to connect the tubes in the order shown in the chart at the
right and matched to their mates labeled on the bowl.
Fix the water lever for the threeway syringe so that it is held down.
There is a device on the bowl to do this.
ADVICE
[5]Operation 7.(1)
Line up and fit gauge
into place
Flushing Switch
W.S. (DH)
Assistant Three-way
Syringe
S.E.
Saliva
Ejector
SC.
Scaler
L.S./H.S.
Micromotor and
Air Turbine Hand pieces
EV.
Vacuum Syringe
W.S. (DR)
Dentist Three-way syringe
Hole in Bowl
(position under auto filler)
Set Up Gauge
(Position by fitting to auto filler)
Flushing Bowl
Flushing Bowl Connections
Instrument
Main Tube Set Up
H.S.
L.S.
SC.
EV.
S.E.
W.S.
Disconnect Hand piece
Disconnect Attachment
Take Tip Off
Take nozzle off