Service manual
Fuse protection schematic for CE Gas units
*
= Option
55
F1 (3,15A) 24V|
Transformer T2
Gas- and Communication pcb
F4 (6,3A)
230V
F2 (6,3A)
230V
Rectifier V8*
F4 (6,3A)
230V
Chamber high limit temp. switch F7/ F9only at 20.x
Motor limiter F6 /F8only at 20.x
Solenoid valve Y12
Reserve relay K23
Supply voltage for control
pcb A2
1,25A 21V
3,15A 10,7V
Supply voltage for chamber
illumination
WaveClean pump M24*
Lidt magnet Y8*
WaveClean pump M16*
Main contactor
K1
Signals for frequency controller V10 / V20
(only at 20x)
Transformer T1
Control pcb A2
Buzzer H13
Cooling fan M7 /M7.1/M7.2
only at 20x
Reset signal ignition box N10 /N20 if error
Er71
(No gas)
only available at gas units
Ignition box N10 (Signal glow electrode)
only at Gas
Ignition box N20 (Signal glow electrode)
only at 20.xGas
F1 (3,15A) 24V|
Transformer T3
Only at 20.x
Power supply for ignition box
(glow electrode) N10
only at gas units
Power supply for gas blower
U10 / U20
only at 20.x gas units
Power supply for ignition box
(glow electrode) N20 only at 20.x gas units
Power supply frequency controller V10
Power supply frequency controller V20
only at 20x
F2 (10A) 230V
F1 (10A) 230V
Part no. of the fuses:
0,8A slow blow = #203720
1,25A slow blow = #203722
2A slow blow = #203474
3,15A slow blow = #203742
6,3A slow blow = #203473
10A slow blow = #203996