![MKN HansDampf 10.1 Service Manual Download Page 38](http://html1.mh-extra.com/html/mkn/hansdampf-10-1/hansdampf-10-1_service-manual_1813327038.webp)
Service manual
CLE
WaveClean™ Test
(Option, only if parameter 170 is set to „on“ in UCF)
With this test the WaveClean System can be tested for leakage and operation (Cartridges will not
dissolve!)
Buttons:
Button
Function
Display
Start WaveClean test
Back to the configuration
menu
After the Start/Stop button is pressed the step starts. The siphon is emptied by the pump M24.
Afterwards the siphon is filled with the drain valve Y12. This step is done twice. Afterwards the pump
M16 and the fan motor M1 /M2 (20.x models) is switched on. The water of the siphon is pumped on
the rotating fan wheel. The procedure can be stopped at any time with the start/stop button.
Afterwards the siphon is emptied and filled twice automatically.
Use this step to test for leakage and operation of the pumps.