MIZUHOSI 2009
NW0504 Rev. D
32
OPTION #3
Use a head halter. Apply the traction cord to the head halter through the
pulley assembly at the head end of the table (see Figure 41). Using
sandbags, apply the desired amount of weight on a weight carrier at the
head end of the table.
Figure 41:
Head halter traction
OPTION #4
Use gardner wells tongs applied to the patient’s skull in accordance with
the manufacturers instructions. Apply traction cord to the tongs through
the pulley assembly at the head end of the table. Using sandbags, apply
the desired amount of weight on a weight carrier at the head end of the
table.
OPTION #5
If the patient is fitted with a cervical halo, it is possible to attach some
halos directly to the Mizuho OSI Cervical Management Base Unit (see
Figure 42). Consult with the halo manufacturer for the availability and use
of halo adapters.
Figure 42:
Halo device attached to the Cervical Management Base Unit
Warning:
The halo, tongs or horseshoe devices should only be used in accordance with the
specific manufacturer’s recommended procedures.
Halo device
Cervical Management Base Unit
Halo adapter
Head halter
Sandbag
weight