MIZUHOSI 2009
NW0504 Rev. D
26
Figure 31:
Mizuho OSI P/N 5300 Lateral Positioner on Radiolucent Imaging Top
6.2.4 The patient may be laterally rotated to a maximum of 25° in each direction. To
do so, use the appropriate table base controls. Do not release the 25° rotation
stop or attempt to rotate the patient in the lateral position more than 25
degrees.
Warning:
The 25° rotational stop must always be engaged when using the Radiolucent
Imaging Top for lateral positioning. Lateral rotation of a patient in the lateral
position to more than 25° may result in the patient being dropped.
NOTE: If increased stability in the lateral position is desired or if more than 25° of
lateral rotation is required for thoracic surgery; we recommend the use of the Mizuho
OSI 5895 Maximum Access™ Lateral Top (see Figure 32). This top is also used for
simultaneous anterior/posterior lumbar surgery. Contact your Mizuho OSI
representative for more information regarding this product.
Lateral
positioners
Imaging top as viewed from above