Installing the NC-2000 Patient Unit
3.8.2 Optional Battery
A rechargeable nickel-cadmium battery (part number NC-PBAT) can be purchased from
Mircom to provide standalone power for the Patient Unit. The battery connects to the battery
terminal CN8 (see Figure 22). To secure the battery, loop a cable tie through the two holes in
the circuit board and then tighten it around the battery.
Figure 22 Battery terminal CN8
3.9 Mounting the Patient Unit
The Patient Unit mounts on the supplied mounting plate, which attaches to a standard 3 gang
PVC box. Note that the Patient Unit is not intended to be used in an Oxygen-Enriched
Environment.
Note:
Do not overtighten the screw terminals.
Attention: The battery alone does not meet the requirements for UL 2560 because
it does not power the dome lights. In order to comply with these
standards, connect the Class 2 power source for the Patient Unit to a
power-limited uninterruptible power source (UPS).
Note:
Prior to mounting the Patient Unit, record its MAC address and location (building,
wing, floor, and room). This information is needed to add the Patient Unit to the
MiCare Central server. Appendix 5: Device Installation Table on page 60 has an
example device installation table that you can use to record this information.
J3
J4
COM
LOW
HIGH
CN16
CN13
CN6
CN2
CN1
CN2
CN8
Battery Terminal
CN8
Holes for
cable tie
Battery -
B