Installing the NC-103 Pull Station
To mount the Pull Station:
1. If you are connecting an input to the Pull Station, run the wires for the input to the single
gang PVC box and then through the wiring hole on the mounting plate (see Figure 29).
2. Attach the mounting plate to the single gang PVC box using two screws (see Figure 29).
Figure 29 Pull Station mounting plate
3. Attach the Pull Station to the mounting plate.
4. Fasten the Pull Station to the mounting plate with a security bolt.
5.5 LED Status Indicator and Beeper
Medium green blinking:
The Pull Station has not joined the network.
Slow green blinking:
The Pull Station is sending a message.
One red blink:
The current operation has failed.
The Pull Station uses two methods to send a message: it first sends a unicast message, and
then a broadcast message if the unicast message fails.
1. Unicast: the Pull Station waits to receive a confirmation from the Nurse Call and
Emergency System Central server that the message was received. The LED blinks red
once if the Pull Station does not receive a reply.
2. Broadcast: the Pull Station does not wait to receive a reply. This means that if the
broadcast message fails, the LED does not blink red. It simply blinks slow green during
the broadcast message, and then stops.
Beep:
The Pull Station beeps when the Emergency button is pressed, when the Pull Station
has found a network, and when the Pull Station receives a reply.
Mounting
Holes
Wiring
Hole
Security
Bolt Hole