25
Installing the NC-2000 Patient Unit
3.5 Paging Input
The paging input terminal, CN2, is located on the keypad circuit board below the main circuit
board (see Figure 16).
Figure 16 External paging input terminal CN2
When connecting to an external paging amplifier, follow these guidelines:
•
Use a dedicated pair of 22AWG wires from each Patient Unit to the paging amplifier.
•
If your paging amplifier outputs 25 V, connect the outputs on the paging amplifier to the
LOW and COM inputs on CN2.
Figure 17 Paging amplifier - 25 V output
Note:
Do not overtighten the screw terminals.
External Paging Input
CN2
J3
J4
COM
LOW
HIGH
CN16
CN13
CN6
CN2
CN1
CN8
i
Paging Amplifier
25V output
22AWG
COM
LOW
COM
C
25V
NC-2000
CN2
COM
LOW
NC-2000
CN2
COM
LOW
NC-2000
CN2
LOW
NC-2000
CN2