50
8.0
Appendix 1: Specifications
8.1
NC-2000 Patient Unit
Standards:
Standard for Safety for Hospital Signaling and Nurse Call Equipment –
UL 1069:2007 Ed.7+R:19 Feb2015
Signal Equipment CSA C22.2#205:2012 Ed.2 +E2015
Emergency Call Systems For Assisted Living And Independent Living
Facilities ANSI/UL 2560:2011 Ed.1 +R:28 May2015
CFR47 FCC Part 15 Subpart C 15.247:2012
CFR47 FCC Part 15 Subpart B:2012
IC RSS-210 Issue 8 December 2010 Annex 8
IC RSS-Gen Issue 3 December 2010
IC ICES-003 Issue 4 February 2004
IC RSS-102 Issue 4 March 2010
CS-03 Part I, issue 9, Amendment 4, December 2010
TIA-968-B-1 (Addendum 1 to TIA-968-B) Approved: June 22, 2012
TIA-968-B-2 (Addendum 2 to TIA-968-B) Approved: January 21, 2015
Power Options:
Input Power:
•
Class 2 power supply 24 VDC, current range 200 mA to 2.5 A
(depending on dome lights - see dome light manufacturer for specific
current requirements)
Paging Wattage Requirements:
•
Minimum 125 mW per unit
Wireless Characteristics:
ZigBee:
•
Frequency: 2.4 GHz
•
Network type: 802.15.4 wireless mesh network
•
Range: 23 m (75 ft) with line of sight
Ultrasonic transmitter:
•
Frequency: 40 kHz
•
Range: 9.1 m (30 ft)
Physical Characteristics:
Faceplate: 221R Lexan
Impact Strength: 260 ft-lbs/in2
Actuator Material: Silicon Rubber
Color: Off White
Weight: 225g (8 oz)
Dimensions: 114 mm x 162 mm x 64 mm (4.5” H x 6.375” W x 2.5” D)
Mounting Centers: 3-9/32”
Standard 3-gang, 2.5” Deep PVC Back Box Required
Operating Temperature: 0º C to +50º C (32º F to 122º F)
Summary of Contents for MiCare NC-103