53
Appendix 1: Specifications
8.4
NC-103 Pull Station
Standards:
Standard for Safety for Hospital Signaling and Nurse Call Equipment –
UL 1069:2007 Ed.7+R:19 Feb2015
Signal Equipment CSA C22.2#205:2012 Ed.2 +E2015
Emergency Call Systems For Assisted Living And Independent Living
Facilities ANSI/UL 2560:2011 Ed.1 +R:28 May2015
CFR47 FCC Part 15 Subpart C 15.247:2012
CFR47 FCC Part 15 Subpart B:2012
IC RSS-210 Issue 8 December 2010 Annex 8
IC RSS-Gen Issue 3 December 2010+Notice 2012-DRS0126
IC ICES-003 Issue 4 February 2004
IC RSS-102 Issue 4 March 2010 updated December 2010
Power Options:
Input Power:
•
2 x 1.5 V Alkaline AA Batteries
•
Nominal Current: 10
μ
A
•
Transmit Current: 35 mA
•
Receive Current: 25 mA
•
Transmit and Receive Frequency: 4 times a day
Wireless Characteristics:
ZigBee:
•
Frequency: 2.4 GHz
•
Network type: 802.15.4 wireless mesh network
•
Range: 23 m (75 ft) with line of sight
Physical Characteristics:
Faceplate: 221R Lexan
Impact Strength: 260 ft-lbs/in2
Color: Off White
Weight: 113 g (4 oz)
Dimensions: 130 mm x 89 mm x 20 mm (5.1” H x 3.5” W x 0.8” D)
Standard 1-gang, 2.5” Shallow PVC Back Box Required
Operating Temperature: 0º C to +50º C (32º F to 122º F)
Input Operating
Specifications:
Wet contact pair, supplied voltage is max 3.6 V
Nominal current of 1.4 mA (external loop closed)
Maximum current 5 mA
Follow manufacturer’s specifications for any attached equipment or wiring.