11. Measurement mode
Selection of measurement mode
S function SF-01
(1) Simple comparison mode
Set the measuring value of SET1 ~ SET8
Details : [11-2.]
Set the comparison value of S0 ~ S4, SET7, SET8
Details : [11-3.]
Simple comparison batch mode
Details of adjusting condition : [11-2-3.]
Simple comparison discharge mode
Details adjusting condition : [11-2-4.]
(2) 5-steps comparator mode
Setting SF-01 [0]
Setting SF-01 [1]
Setting SF-01 [0]
Setting SF-01 [1]
Selection of control mode
S function SF-02
Details : [11-1. Selection method of measurement mode]
Details : [9-2. Function of S function data]
10. Various operations by S function data
CSD-892-73 consists of two(2) measurement mode as [Simple comparison mode] and [5
steps comparator mode], and furthermore, [Simple comparison mode] includes [Batch mode]
and [Discharge mode]. The comparison result of the set value and the measuring value can
be output by selecting either of mode, and also setting the necessary parameter. The outside
equipment such as PLC and PC can be controlled by allocating the function to the external
control output ([OUTPUT1] ~ [OUTPUT5]) of a lower terminal block. Please refer to the
followings for the selection method and the setting method of each mode.
The outline of each mode is as follows:
(1) Simple calibration mode
Whenever the result of comparing the set value with the measuring value meets the
condition of judgement, the output is always ON.
There are two(2) control mode in the simple comparison mode, and either [Simple
comparison batch mode] or [Simple comparison discharge mode] is selected.
Please refer to [11-2.] for the details of comparison and judgement condition.
(2) 5-steps comparator mode
The result of comparing the set value with measuring value is output.
After the measuring value is compared with one of the set values, ON/OFF of the output is
decided on the condition of over and under.
The comparison value can be set up to
fi
ve from S0 ~ S4.
Please refer to each item in [11-3.] for the details of operational setting.