V
5. Conformity standard
This instrument has suited the following standard.
EN61326-1
:
2006
"Electrical equipment for measurement, control, and laboratory use
-
EMC requirements"
"Immunity test requirements for equipment intended for use in
industrial locations"
•
Installation place
Please set up in the control board and the operation panel where EMC solution is
given.
•
Power supply
Please set up the power supply of DC24 V in the control board and operation
panel where EMC solution is also given.
The power supply of DC24 V must use "CE applied product".
•
Cable
Use the shielded cable other than the power cable.
•
Shield processing
Please connect the shield cable of load cell with E terminal in upper terminal
block.
•
Please connect the shield cable only for CC-Link interface with SLD terminal in
the connector for CC-Link interface.
Please process the shield to the cable for CC-Link and the external I/O surely by
the conduit piping including the control board.
•
Grounding
The ground shall apply the individual ground by using the F.G. terminal in lower
terminal block.
Please make the single ground of the instrument, or of the storage case after
connecting with the storage case when you store only the DC24 V power circuit
and the instrument in the storage case.
Please give the earth as a single earth when you store the device other than the
instrument in the storage case.
Caution
•
At the time of shipment from the factory, the instrument has been plated with a
clear sheet on the panel sheet for protective purpose. In case of application, use
the instrument after removing the clear sheet
fi
rst.
•
Do not push the panel sheet on the instrument with the excessive strong force nor
push it with sharp edge object such as a driver.
•
If neglected, it may cause damage to the panel switch and even have the
possibility of damage to resist to environments or operational performances.
•
When the USB interface is used, the applied CE standard is excluded.
•
Please observe the following conditions strictly when this instrument suits the below. If
neglected, it may not conform to the above standard.