OL
This lights on with
when the load display exceeds over [(+ Weighing
capacity) + 9D] or [+110 % of the weighing value]. (By the setting of C function
CF-03.)
- OL
This lights on with
when the load display exceeds over [(- Weighing
capacity) - 9D] or [-110 % of the weighing value]. (By the setting of C function
CF-03.)
OVF
Plus over of A/D conversion input. This lights on with
.
- OVF
Minus over of A/D conversion input. This lights on with
.
All light on
-
(When C function CF-15 [Power on ZERO operation is valid.)
All the segments light on when the stability is not detected at turning on the
power. By pushing [FUNC.] key in the condition of all lighting on, the display
becomes the load display compulsorily.
HOLD
This lights on when the power turns on in the condition of HOLD signal ON.
By releasing from the HOLD signal, the display becomes the load display.