SP-H error
When the read load cell output value is over than 3.1 mV/V and exceeds over SPAN
adjustment range at plus side, the following code is set.
Condition
Error code
Error assistance code
SP-H error
4
3
Error that cannot be executed
The following code is set when executing by other than the calibration mode, or not setting it
according to the
fl
ow chart in [14-6.].
Condition
Error code
Error assistance code
Error that cannot be executed.
99
2
⑬
End of calibration
The calibration data of the ZERO/SPAN, etc., that is registered temporary is stored by writing
the data in the table below.
Command No.
(RWwnE) : 4 stations are occupied.
(RWwn6) : 2 stations are occupied.
General data area
(RWwnC~RWwnD)
(RWwn4~RWwn5)
9099
0
Error that cannot be executed
The following code is set when executing by other than the calibration mode, or not setting
it according to the
fl
ow chart in [14-6.].
Condition
Error code
Error assistance code
Error that cannot be executed.
99
2
⑭
Calibration forced
fi
nish
During the calibration procedure, the calibration is discontinued without storing the calibration
data that is registered temporarily by writing the data in the table below, and you can return
to the measurement mode.
When the data other than [0] are written in the general data area, the signal becomes an
error outside a set range. (Error code : 99, Error assistance code : 1)
Command No.
(RWwnE) : 4 stations are occupied.
(RWwn6) : 2 stations are occupied.
General data area
(RWwnC~RWwnD)
(RWwn4~RWwn5)
9000
0
Error that cannot be executed
The following code is set when executing by other than the calibration mode.
Condition
Error code
Error assistance code
Error that cannot be executed.
99
2
107
C
C
-Li