7-7
Module
Parameter
Low alarm limit
High alarm limit
Auto alarm limits range
(Cuff Type: Big) NIBP-D
Dia×0.68 + 6mmHg
Dia×0.86 + 32mmHg
25 to 210
NIBP-M
Mean×0.68 + 8mmHg
Mean×0.86 + 35mmHg
30 to 230
Temp
T1
(T1 – 0.5) °C
(T1 + 0.5) °C
1 to 49 °C
T2
(T2 – 0.5) °C
(T2 + 0.5) °C
1 to 49 °C
TD
Same as the default alarm
limit
Same as the default alarm limit Same as the measurement range
IBP
:
ART/
Ao/
UAP/
BAP/
FAP/
LV/
P1-P4
(Arterial
pressure)
IBP-S
(SYS – 15) or 45mmHg
(whichever is greater)
(SYS + 15) or 105mmHg
(whichever is smaller)
35 to 115
IBP-D
(Dia – 15) or 20mmHg
(whichever is greater)
(Dia + 15) or 80mmHg
(whichever is smaller)
20 to 90
IBP-M
(Mean – 15) or 35mmHg
(whichever is greater)
(Mean + 15) or 95mmHg
(whichever is smaller)
25 to 105
PA
IBP-S
SYS × 0.75
SYS × 1.25
3 to 120mmHg
IBP-D
Dia × 0.75
Dia × 1.25
IBP-M
Mean × 0.75
Mean × 1.25
IBP:
CVP/
ICP/
LAP/
RAP/
UVP/
P1-P4
(Venous
pressure)
IBP-M
Mean × 0.75
Mean × 1.25
3 to 40mmHg
CO
2
EtCO
2
0 to 32mmHg:
remains the same
0 to 32mmHg:
remains the same
Same as the measurement range
32 to 35mmHg:
29mmHg
32 to 35mmHg:
41mmHg
35 to 45mmHg:
(etCO
2
-6) mmHg
35 to 45mmHg:
(etCO
2
+6) mmHg
45 to 48mmHg:39 mmHg
45 to 48mmHg:51 mmHg
>48mmHg: remains the
same
>48mmHg: remains the same
FiCO
2
N/A
Same as the default alarm limit Same as the measurement range
awRR
(awRR – 10) or 30 rpm
(whichever is greater)
(awRR+25) or 85 rpm
(whichever is smaller)
10 to 90
Summary of Contents for iMEC8 Vet
Page 1: ...iMEC8 Vet Portable Multi Parameter Veterinary Monitor Operator s Manual ...
Page 2: ......
Page 16: ...X F 1 Symbols F 1 F 2 Abbreviations F 2 ...
Page 54: ...6 6 FOR YOUR NOTES ...
Page 94: ...11 8 FOR YOUR NOTES ...
Page 170: ...A 14 FOR YOUR NOTES ...
Page 176: ...B 6 FOR YOUR NOTES ...
Page 201: ......
Page 202: ...PN 046 008143 00 3 0 ...