A-7
(not available in USA)
Waveform
4ah - range:
4a - range:
4ad - range:
Waveform 4bh - range:
4b - range:
4bd - range:
< 11 s
< 11 s
< 11 s
< 11 s
< 11 s
< 11 s
Tall T-wave rejection capability
When the test is performed based on part 4.1.2.1 c)of ANSI/AAMI EC 13-2002, the
heart rate meter will reject all 100 ms QRS complexes with less than 1.2 mV of
amplitude, and T waves with T-wave interval of 180 ms and those with Q-T interval
of 350 ms.
Arrhythmia Analysis Classifications
Asystole, VFib/VTac, Vtac, Vent. Brady, Extreme Tachy, Extreme Brady, PVC, Couplet,
Bigeminy, Trigeminy, R on T, VT>2, PVCs, Tachy, Brady, Missed Beats, Vent. Rhythm,
PNP, PNC, Multif. PVC, Nonsus. Vtac, Pause, Irr. Rhythm
ST Segment Analysis (Not available in USA)
Measurement range
-2.0 to 2.0 mV RTI
Accuracy
-0.8 to 0.8 mV:
Beyond this range:
±0.02 mV or ±10%, whichever is greater.
Not specified.
Refreshing rate
10 s
Resolution
0.01 mV
A.7.2
Resp
Technique
Trans-thoracic impedance
Lead
Options are lead I and II. The default is lead II.
Respiration excitation waveform
<300 µA RMS, 62.8 kHz (±10%)
Baseline impedance range
200 to 2500Ω (using an ECG cable with 1kΩ resistance)
Bandwidth
0.2 to 2 Hz (-3 dB)
Sweep speed
3 mm/s, 6.25 mm/s, 12.5 mm/s, 25 mm/s, or 50 mm/s
Accuracy: ±10%
Respiration Rate
Measurement range
0 to 150 rpm
Resolution
1 rpm
Accuracy
7 to 150 rpm:
0 to 6 rpm:
±2 rpm or ±2%, whichever is greater
Not specified.
Apnea alarm time
10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s
Alarm limit
Range (rpm)
Step (rpm)
RR High
(low limit + 2) to 150
1
RR Low
0 to (high limit – 2)
Summary of Contents for iMEC8 Vet
Page 1: ...iMEC8 Vet Portable Multi Parameter Veterinary Monitor Operator s Manual ...
Page 2: ......
Page 16: ...X F 1 Symbols F 1 F 2 Abbreviations F 2 ...
Page 54: ...6 6 FOR YOUR NOTES ...
Page 94: ...11 8 FOR YOUR NOTES ...
Page 170: ...A 14 FOR YOUR NOTES ...
Page 176: ...B 6 FOR YOUR NOTES ...
Page 201: ......
Page 202: ...PN 046 008143 00 3 0 ...