D-5
Measurement
Alarm Message
L
I
Cause and Solution
CO
2
Low Barometric
Press.
L
C
sources that affect the ambient pressure. Restart the monitor.
CO
2
FilterLine
Occluded
L
C
The airway or watertrap was occluded. Check the airway and
remove the occlusion.
CO
2
No Watertrap
L
B
Check the watertrap connections.
CO
2
Check Adapter
L
A
There is a problem with the airway adapter. Check, clean or
replace the adapter.
CO
2
FilterLine Err
L
C
Check if there is a leak in the CO
2
sample line or the CO
2
sample line has been occluded.
CO
2
Zero Failed
L
A
Check the CO
2
connections. After the sensor’s temperature
becomes stabilized, perform a zero calibration again.
CO
2
System Err
L
A
Re-plug the module or restart the monitor.
CO
2
Check Cal.
L
C
Perform a calibration.
CO
2
Check Airway
L
C
An error occurred to the airway.
CO
2
No Filterline
L
A
Make sure that the filterline is connected.
CO
2
No Sensor
L
A
Make sure that the sensor is connected.
CO
2
Main Board Err
H
C
There is a problem with the CO
2
module. Re-plug the module
or restart the monitor.
CO
2
Checking Sensor
L
C
CO
2
Replace
Scrubber&Pump
L
C
CO
2
15V Overrange
H
C
CO
2
Hardware Err
H
C
Power
12V Too High
H
C
There is a problem with the system power supply. Restart the
monitor.
12V Too Low
H
C
5V Too High
H
C
5V Too Low
H
C
3.3V Too High
H
C
3.3V Too Low
H
C
Battery Too Low
H
C
Connect the monitor to an AC power source and allow the
batteries to charge.
Different Battery
Voltages
M
C
The two batteries have different charge capacity, or the
batteries unspecified have been used, or there is a problem
with the batteries. Make sure that correct batteries are used
and the batteries are not damaged, or replace the batteries.
Battery Power
Overload
H
C
The power consumption of the equipment is too high. Power
the monitor using an AC power source.
RT Clock Not Exist
H
C
Contact your service personnel.
Recorder
Recorder Init Err N
L
A
Restart the monitor.
N is within 1 to 8.
Recorder SelfTest Err
L
A
Stop the recording and restart the monitor.
Recorder Comm Err
L
A
Recorder S. Comm
Err
L
A
Summary of Contents for iMEC8 Vet
Page 1: ...iMEC8 Vet Portable Multi Parameter Veterinary Monitor Operator s Manual ...
Page 2: ......
Page 16: ...X F 1 Symbols F 1 F 2 Abbreviations F 2 ...
Page 54: ...6 6 FOR YOUR NOTES ...
Page 94: ...11 8 FOR YOUR NOTES ...
Page 170: ...A 14 FOR YOUR NOTES ...
Page 176: ...B 6 FOR YOUR NOTES ...
Page 201: ......
Page 202: ...PN 046 008143 00 3 0 ...