15-5
15.3.6
Setting the Apnea Alarm Delay
In the [
CO2 Setup
] menu, select [
Apnea Delay
] and then select the appropriate setting. The monitor will alarm if the
patient has stopped breathing for longer than the preset apnea time. The [
Apnea Delay
] of Resp and CO
2
module keeps
consistent with each other.
WARNING
The respiration measurement does not recognize the cause of apneas. It only indicates an alarm if no breath
is detected when a pre-adjusted time has elapsed since the last detected breath. Therefore, it cannot be
used for diagnostic purpose.
15.3.7
Choosing a Time Interval for Peak-Picking
For microstream CO
2
modules, you can select a time interval for picking the highest CO
2
as the EtCO
2
and the lowest as
the FiCO
2
.
In the [
CO
2
Setup
] menu, select [
Max Hold
] and toggle between [
Single Breath
], [
10 s
], [
20 s
] and [
30 s
] (for
microstream CO
2
module only).
[
Single Breath
]: EtCO
2
and FiCO
2
are calculated for every breath.
[
10 s
] or [
20 s
], or [
30 s
]: EtCO
2
and FiCO
2
are calculated using 10, 20 or 30 seconds of data.
15.3.8
Setting the Flow Rate
For the sidestream CO
2
module, you can change the sampling rate of respiratory gas in the patient’s airway by setting
the flow rate. To set the flow rate, enter the [
CO2 Setup
] menu and select an appropriate setting from [
Flow Rate
].
WARNING
Please consider the patient’s actual bearing capability and select the appropriate flow rate when setting the
flow rate.
15.3.9
Setting up the CO
2
Wave
In the [
CO
2
Setup
] menu, you can:
Select [
Wave Type
] and toggle between [
Draw
] and [
Fill
]:
[
Draw
]: The CO
2
wave is displayed as a curved line.
[
Fill
]: The CO
2
wave is displayed as a filled area.
Select [
Sweep
] and then select the appropriate setting. The faster the wave sweeps, the wider the wave is.
Change the size of the CO
2
waveform by adjusting the wave [
Scale
].
Summary of Contents for iMEC8 Vet
Page 1: ...iMEC8 Vet Portable Multi Parameter Veterinary Monitor Operator s Manual ...
Page 2: ......
Page 16: ...X F 1 Symbols F 1 F 2 Abbreviations F 2 ...
Page 54: ...6 6 FOR YOUR NOTES ...
Page 94: ...11 8 FOR YOUR NOTES ...
Page 170: ...A 14 FOR YOUR NOTES ...
Page 176: ...B 6 FOR YOUR NOTES ...
Page 201: ......
Page 202: ...PN 046 008143 00 3 0 ...