B-4
Recommended separation distances between portable and mobile RF communications equipment and the device
The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the device as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter (W)
Separation distance in meters (m) according to frequency of the transmitter
150 kHz
~
80 MHz
P
d
2
.
1
=
80 MHz
~
800 MHz
P
d
2
.
1
=
(Resp:
P
d
5
.
3
=
)
800 MHz
~
2.5 GHz
P
d
3
.
2
=
( Resp:
P
d
7
=
)
0.01
0.12
0.12 (0.35)
0.23 (0.70)
0.1
0.38
0.38 (1.11)
0.73 (2.22)
1
1.20
1.20 (3.50)
2.30 (7.00)
10
3.80
3.80 (11.07)
7.30 (22.14)
100
12.00
12.00 (35.00)
23.00 (70.00)
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m)
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
B.2
Radio Regulatory Compliance
RF parameters
Item
Description
IEEE 802.11b
IEEE 802.11g
IEEE 802.11n
Operating Frequency Band
(MHz)
2412 - 2472
2412 - 2472
2412 - 2472
Modulation
DSSS and CCK
OFDM
OFDM
Transmitter Output Power
(dBm)
<20
<20
<20
The device configured with the WM1010BGN wireless module complies with part 15 of the FCC Rules. Operation is
subject to the condition that this device does not cause harmful interference.
Operation of this equipment requires the prior coordination with a frequency coordinator designated by the FCC for the
Wireless Medical Telemetry Service.
This Wi-Fi device complies with Canadian ICES-001. Cet appareil ISM est conforme a la norme NMB-001 du Canada.
Summary of Contents for iMEC8 Vet
Page 1: ...iMEC8 Vet Portable Multi Parameter Veterinary Monitor Operator s Manual ...
Page 2: ......
Page 16: ...X F 1 Symbols F 1 F 2 Abbreviations F 2 ...
Page 54: ...6 6 FOR YOUR NOTES ...
Page 94: ...11 8 FOR YOUR NOTES ...
Page 170: ...A 14 FOR YOUR NOTES ...
Page 176: ...B 6 FOR YOUR NOTES ...
Page 201: ......
Page 202: ...PN 046 008143 00 3 0 ...