15-7
15.8
Troubleshooting the Sidestream CO
2
Sampling System
When the sampling system of the sidestream CO
2
module works incorrectly, check if the sampling line is kinked. If not,
remove it from the watertrap. If the monitor gives a message indicating the airway still works incorrectly, it indicates that
the watertrap must have been blocked, and you should replace with a new one. Otherwise, you can determine that the
sampling line must have been blocked. Replace with a new sampling line.
15.9
Removing Exhaust Gases from the System
WARNING
Anesthetics: When using the Sidestream or Microstream CO
2
measurement on patients who are receiving or
have recently received anesthetics, connect the outlet to a scavenging system, or to the anesthesia
machine/ventilator, to avoid exposing medical staff to anesthetics.
To remove the sample gas to a scavenging system, connect an exhaust tube to the gas outlet connector of the module.
15.10
Zeroing the Sensor
The zero calibration eliminates the effect of baseline drift during CO
2
measurement exerted on the readings and
therefore maintains the accuracy of the CO
2
measurements.
15.10.1
For Sidestream and Microstream CO
2
Modules
For sidestream and microstream CO
2
modules, a zero calibration is carried out automatically when necessary. You can
also start a manual zero calibration if necessary.
To manually start a zero calibration:
1.
Select [
Main Menu
]
→
[
Maintenance
>>]
→
[
User Maintenance
>>]
→
enter the required password
→
[
Maintain
CO2
>>].
2.
Select [
Calibrate CO2
>>]
→
[
Start Zero Cal.
].
Disconnecting the patient airway is not required when performing a zero calibration.
15.11
Calibrating the Sensor
For sidestream or microstream CO
2
modules, a calibration should be performed once every year or when the readings go
far beyond the range. For details, refer to the chapter
Summary of Contents for iMEC8 Vet
Page 1: ...iMEC8 Vet Portable Multi Parameter Veterinary Monitor Operator s Manual ...
Page 2: ......
Page 16: ...X F 1 Symbols F 1 F 2 Abbreviations F 2 ...
Page 54: ...6 6 FOR YOUR NOTES ...
Page 94: ...11 8 FOR YOUR NOTES ...
Page 170: ...A 14 FOR YOUR NOTES ...
Page 176: ...B 6 FOR YOUR NOTES ...
Page 201: ......
Page 202: ...PN 046 008143 00 3 0 ...