17-2
A time indicating your current position is displayed above the parameter area. Numeric measurement values
corresponding to the cursor location change as the cursor is moved. The measurement value that triggered high
level alarm has red background. The one that triggered medium/low level alarm has yellow background.
By selecting
or
beside [
Event
], you can position the cursor to different event time.
By selecting the [
Record
] button, you can print out the currently displayed graphic trends by the recorder.
17.3
Reviewing Tabular Trends
In the [
Review
] menu, select [
Tabular Trends
] to access the Tabular Trends window.
Events are marked with colors in window’s top area. Red represents high level alarm event. Yellow represents
medium/low level alarm event. Green represents manual event.
In this review window:
Select [
Trend Group
] and you can select a trend group for viewing in the popup menu. If [
Custom 1
] or [
Custom 2
]
is selected, you can further select [
Define Trend Group
]. Then you can select the parameters for viewing in the
popup menu.
You can change the resolution of the trend data by selecting [
Interval
] and then selecting the appropriate setting:
[
5 s
] or [
30 s
]: select to view up to 4 hours of tabular trends at 5- or 30-second resolution.
[
1 min
], [
5 min
], [
10 min
], [
15 min
], [
30 min
], [
1 h
], [
2 h
] or [
3 h
]: select to view up to 120 hours of tabular
trends at your selected resolution.
[
NIBP
]: select to view the tabular trends when NIBP measurements were acquired.
To browse the tabular trends, you can either:
Select
or
beside [
Scroll
] to drag the scrollbar left or right to navigate through the trend database, or
Select
or
to scroll left or right to navigate through the trend database.
The measurement value that triggered high level alarm has red background. The one that triggered medium/low level
alarm has yellow background.
By selecting
or
beside [
Event
], you can position the cursor to different event time.
By selecting the [
Record
] button, you can access the [
Record Setup
] menu and set the start and end time of the
tabular trends you want to record. This feature is not available when reviewing a history patient. By further
selecting [
Record
], you can print out the currently displayed tabular trends by the recorder.
17.4
Events
17.4.1
Marking Events
During monitoring, some events may exert effects on the patient and as a result change the waveforms or numerics
displayed on the monitor. To help analysing the waveforms or numerics at that time, you can mark these events.
Select [
Main Menu
]
→
[
Mark Event >>
]. In the popup menu, you can select the waves to be stored when a manual event
is triggered. You can select [
Trigger Manual Event
] from the [
Mark Event
] menu or the [
Manual Event
] QuickKey to
trigger a manual event and store it at the same time.
Summary of Contents for iMEC8 Vet
Page 1: ...iMEC8 Vet Portable Multi Parameter Veterinary Monitor Operator s Manual ...