15-2
15.2
Measuring CO
2
15.2.1
Making a Sidestream CO
2
Measurement
1. Attach the watertrap to the monitor and then connect the CO
2
components as shown below.
2. The CO
2
module needs time to warm up to reach the operating temperature. The message [
CO2 Sensor Warmup
]
is displayed during warm-up. If you perform CO
2
measurements during warm-up, the measurement accuracy may
be compromised.
3. After warm-up is finished, you can perform CO
2
measurements.
CAUTION
The watertrap collects water drops condensed in the sampling line and therefore prevents them from
entering the module. If the collected water reaches a certain amount , you should drain it to avoid blocking
the airway.
The watertrap has a filter preventing bacterium, water and secretions from entering the module. After
long-term use, dust or other substances may compromise the performance of the filter or even block the
airway. In this case, replace the watertrap. It is recommended to replace the watertrap once a month, or
when the watertrap is found leaky, damaged or contaminated.
NOTE
To extend the lifetime of the watertrap and module, disconnect the watertrap and set the operating mode
to standby mode when CO
2
monitoring is not required.
Watertrap
Watertrap fixer
Sampling line
Summary of Contents for iMEC8 Vet
Page 1: ...iMEC8 Vet Portable Multi Parameter Veterinary Monitor Operator s Manual ...
Page 2: ......
Page 16: ...X F 1 Symbols F 1 F 2 Abbreviations F 2 ...
Page 54: ...6 6 FOR YOUR NOTES ...
Page 94: ...11 8 FOR YOUR NOTES ...
Page 170: ...A 14 FOR YOUR NOTES ...
Page 176: ...B 6 FOR YOUR NOTES ...
Page 201: ......
Page 202: ...PN 046 008143 00 3 0 ...