Maintenance
WARNING
Failure on the part of the responsible individual hospital or institution employing the use of this equipment
to implement a satisfactory maintenance schedule may cause undue equipment failure and possible health
hazards.
The safety checks or maintenance involving any disassembly of the equipment should be performed by
professional servicing personnel. Otherwise, undue equipment failure and possible health hazards could
result.
If you discover a problem with any of the equipment, contact your service personnel or Mindray.
23.1
Regular Inspection
Before the first use, after your monitor has been used for 6 to 12 months, or whenever your monitor is repaired or
upgraded, a thorough inspection should be performed by qualified service personnel to ensure the reliability.
Follow these guidelines when inspecting the equipment:
Make sure that the environment and power supply meet the requirements.
Inspect the equipment and its accessories for mechanical damage.
Inspect all power cords for damage, and make sure that their insulation is in good condition.
Make sure that only specified accessories are applied.
Inspect if the alarm system functions correctly.
Make sure that the recorder functions correctly and the recorder paper meets the requirements.
Make sure that the batteries meet the performance requirements.
Make sure that the monitor is in good working condition.
In case of any damage or abnormity, do not use the monitor. Contact the hospital’s biomedical engineers or your service
personnel immediately.
Summary of Contents for iMEC8 Vet
Page 1: ...iMEC8 Vet Portable Multi Parameter Veterinary Monitor Operator s Manual ...
Page 2: ......
Page 16: ...X F 1 Symbols F 1 F 2 Abbreviations F 2 ...
Page 54: ...6 6 FOR YOUR NOTES ...
Page 94: ...11 8 FOR YOUR NOTES ...
Page 170: ...A 14 FOR YOUR NOTES ...
Page 176: ...B 6 FOR YOUR NOTES ...
Page 201: ......
Page 202: ...PN 046 008143 00 3 0 ...