20-2
20.3.2
Transferring Data by Different Means
NOTE
Never enter the data transfer mode when the monitor is in normal operation or performs monitoring. You
must re-start the monitor to exit the data transfer mode.
Transfer data via a crossover network cable
Before transferring data using a crossover network cable, do as follows:
1. Connect one end of the crossover network cable to the monitor and the other end to the PC.
2. Set the IP address of the PC. This IP address must be in the same network segment with that of the monitor.
3. Make sure that the data export system is active on the PC.
Then, follow this procedure to transfer data:
1. Select [
Main Menu
]
→
[
Patient Data >>
]
→
[
Transfer Data
].
2. Select [
Yes
] from the popup message box.
3. Input the IP address already set on the PC.
4. Select [
Start
] to start transferring data.
Transfer data within a LAN
Before transferring data within a LAN, do as follows:
1. Connect the monitor and the intended PC into the same LAN and acquire the PC’s IP address.
2. Make sure that the data export system is active on the PC.
Follow the same procedure as via a crossover network cable to transfer data.
20.4
Nurse Call
The monitor provides a nurse call connector to output nurse call signal when a userdefined alarm occurs. To obtain
nurse call signal, use the nurse call cable (
PN: 009-003116-00
) we supply to connect the hospital nurse call system to the
nurse call connector of the monitor and then follow this procedure:
1. Select [
Main Menu
]
→
[
Maintenance >>
]
→
[
User Maintenance >>
]
→
enter the required password.
2. Select [
Others >>
] to access the [
Others
] menu.
3. Select [
Nurse Call Setup >>
] to change the nurse call settings as follows:
Select [
Signal Type
] and toggle between [
Pulse
] and [
Continuous
].
[
Pulse
]: the nurse call signals are pulse signals and each pulse lasts 1 second. When multiple alarms occur
simultaneously, only one pulse signal is outputted. If an alarm occurs but the previous one is not cleared yet, a
new pulse signal will also be outputted.
[
Continuous
]: the nurse call signal lasts until the alarm ends, i.e. the duration of a nurse call signal equals to
that of the alarm condition.
Summary of Contents for iMEC8 Vet
Page 1: ...iMEC8 Vet Portable Multi Parameter Veterinary Monitor Operator s Manual ...
Page 2: ......
Page 16: ...X F 1 Symbols F 1 F 2 Abbreviations F 2 ...
Page 54: ...6 6 FOR YOUR NOTES ...
Page 94: ...11 8 FOR YOUR NOTES ...
Page 170: ...A 14 FOR YOUR NOTES ...
Page 176: ...B 6 FOR YOUR NOTES ...
Page 201: ......
Page 202: ...PN 046 008143 00 3 0 ...