EN
| INSTRUCTIONS FOR USE
proGAV
®
Tools
Item no.
Figure
9
FV409T
3.01.02 SCOPE OF DELIVERY
Box content
(item no. FV404T)
Number
proGAV Adjustment Tool
1
proGAV Verification Tool
1
proGAV Masterdisc
1
proGAV Verification Compass
1
proGAV Instrument Guide
1
proGAV Tools
instructions for use
1
Box content
(item no. FV409T)
Number
proGAV Checkmate
1
proGAV Tools
instructions for use
1
Box content
(item no. FV499T)
Number
proGAV Adjustment Tool
1
proGAV Verification Tool
1
proGAV Masterdisc
1
proGAV Verification Compass
1
proGAV Adjustment Disc
, small 1
proGAV Adjustment Disc
,
medium
1
proGAV Adjustment Disc
, large
1
proGAV Instrument Guide
1
proGAV Tools
instructions for use
1
3.01.03 STERILITY
With the exception of the
proGAV Checkmate
,
proGAV Tools
cannot be sterilised, only disin-
fected.
WARNING
If the packaging or the product is damaged,
the products must not be used under any cir-
cumstances.
3.01.04 REPEATED USE AND
RE-STERILISATION
The
proGAV Checkmate
must be sterilised in
double sterile packaging in an autoclave (steam
sterilisation, fractionated vacuum method) at
134 °C and 5 minutes retention time.
3.01.05 OPERATING CONDITIONS
The
proGAV Tools
are to be used at room tem-
perature in healthcare facilities, such as hospi-
tals, doctor’s surgeries and rehabilitation facili-
ties.
3.01.06 PRODUCT LIFE
These medical devices are designed to oper-
ate accurately and reliably over long periods of
time. The expected product life of the
proGAV
Tools
is three years after first use, provided that
the product is exposed to normal usage con-
ditions and is properly maintained (see chapter
“3.04 Use of the product”).
The instruments may be used beyond this time
if they are in perfect working condition. How-
ever, we cannot guarantee that these medical
devices will require replacement for medical or
technical reasons.
3.01.07 PRODUCT CONFORMITY
The product meets current regulatory require-
ments.
Translations of these instructions for use into
additional languages can be found on our web-
site:
https://www.miethke.com/downloads/
3.02 IMPORTANT SAFETY INFORMATION
3.02.01 SAFETY INSTRUCTIONS
Important!
Read all safety instructions care-
fully before using the product. Follow the safety
instructions in order to avoid injuries and life-
threatening situations.
WARNING
If the sterile packaging of the
Checkmate
is
damaged, the product must not be used.
The
Checkmate
must not be used after the
expiry date and must be re-sterilised as
described under chapter 3.01.04.
26