7

BP A110

EN

- direct sunlight

- heat and cold

The cuffs are sensitive and must be handled with care.

Only pump up the cuff once fitted.

The function of this device may be compromised when used 

close to strong electromagnetic fields such as mobile phones or 

radio installations and we recommend a distance of at least 

1 m. In cases where you suspect this to be unavoidable, please 

verify if the device is working properly before use.

Do not use this device if you think it is damaged or notice 

anything unusual.

Never open this device.

If the device is not going to be used for a prolonged period the 

batteries should be removed.

Read the additional safety information provided within the indi-

vidual sections of this instruction manual.

Ensure that children do not use this device unsupervised; 

some parts are small enough to be swallowed. Be aware of 

the risk of strangulation in case this device is supplied with 

cables or tubes.

Device care

Clean the device only with a soft, dry cloth.

Cleaning the cuff

Carefully remove spots on the cuff with a damp cloth and soapsuds.

WARNING: 

Do not wash the cuff in a washing machine or 

dishwasher!

Accuracy test

We recommend this device is tested for accuracy every 2 years or 

after mechanical impact (e.g. being dropped). Please contact your 

local Microlife-Service to arrange the test (see foreword).

Disposal

Batteries and electronic devices must be disposed of in 

accordance with the locally applicable regulations, not with 

domestic waste. 

12. Guarantee

This device is covered by a 

5 year guarantee

 from the date of 

purchase. The guarantee is valid only on presentation of the guar-

antee card completed by the dealer (see back) confirming date of 

purchase or the receipt.

Batteries, cuff and parts that become worn with use are not 

included.

Opening or altering the device invalidates the guarantee.

The guarantee does not cover damage caused by improper 

handling, discharged batteries, accidents or non-compliance 

with the operating instructions.

Please contact your local Microlife-Service (see foreword).

13. Technical Specifications

This device complies with the requirements of the Medical Device 

Directive 93/42/EEC.

Technical alterations reserved.
Date of production: first four digits of the serial number of the 

device. First and second digit – the week of the year / third and 

fourth digit – the year of production.

Operating conditions:

10 - 40 °C / 50 - 104 °F

15 - 95 % relative maximum humidity

Storage conditions:

-20 - +55 °C / -4 - +131 °F

15 - 95 % relative maximum humidity

Weight:

560 g (including batteries)

Dimensions:

152 x 92 x 42 mm

Measuring procedure:

oscillometric, corresponding to Korotkoff 

method: Phase I systolic, Phase V diastolic

Measurement range:

20 - 280 mmHg – blood pressure

40 - 200 beats per minute – pulse

Cuff pressure display 

range:

0 - 299 mmHg

Resolution:

1 mmHg

Static accuracy:

pressure within ± 3 mmHg

Pulse accuracy:

± 5 % of the readout value

Voltage source:

4 x 1.5 V alkaline batteries; size AA

Mains adapter DC 6V, 600 mA (optional)

Battery lifetime:

approx. 920 measurements 

(using new batteries)

Included accessories:

blood pressure monitor, cuff, bag, 4 x AA 

batteries, instruction manual, guarantee 

card, blood pressure diary

IP Class:

IP20

Reference to 

standards:

EN 1060-1 /-3 /-4; IEC 60601-1;

IEC 60601-1-2 (EMC); IEC 60601-1-11

Expected service life:

Device: 5 years or 10000 measurements

Accessories: 2 years

Summary of Contents for BP A110

Page 1: ...ww microlife com Asia Microlife Corporation 9F 431 RuiGang Road NeiHu Taipei 11492 Taiwan R O C Tel 886 2 8797 1288 Fax 886 2 8797 1283 Email service microlife com tw www microlife com North Central S...

Page 2: ...5 BP A110 UA o Microlife 12 5 o Microlife 13 93 42 EEC 10 40 C 15 95 20 55 C 15 95 560 152 x 92 x 42 I V 20 280 40 200 0 299 1 3 5 4 x 1 5 AA 6 600 920 AA 4 IP IP20 EN 1060 1 3 4 IEC 60601 1 IEC 60601...

Page 3: ...etects patients with AFIB at 97 100 certainty 1 2 Please read through these instructions carefully so that you under stand all functions and safety information We want you to be happy with your Microl...

Page 4: ...h blood pressure values can damage your health and must be treated by your doctor Always discuss your values with your doctor and tell him her if you have noticed anything unusual or feel unsure Never...

Page 5: ...tery compartment AT Setting the date and time 1 After the new batteries are fitted the year number flashes in the display You can set the year by pressing the M button 2 To confirm and then set the mo...

Page 6: ...7 The cuff will now pump up automatically Relax do not move and do not tense your arm muscles until the measurement result is displayed Breathe normally and do not talk 8 When the correct pressure is...

Page 7: ...y capacity of 200 is not exceeded When the memory is full the old values are automatically overwritten with new ones Values should be evaluated by a doctor before the memory capacity is reached otherw...

Page 8: ...fety and protection This device may only be used for the purposes described in these instructions The manufacturer cannot be held liable for damage caused by incorrect application This device comprise...

Page 9: ...ase The guarantee is valid only on presentation of the guar antee card completed by the dealer see back confirming date of purchase or the receipt Batteries cuff and parts that become worn with use ar...

Page 10: ...riakidis M Diagnostic accuracy of a home blood pressure monitor to detect atrial fibrillation J Hum Hyperten 2009 1 5 2 Wiesel J Fitzig L Herschman Y Messineo FC Detection of Atrial Fibrillation Using...

Page 11: ...9 BP A110 RU 1 2 AFIB AFIB AFIB Microlife MAM 3 MAM MAM 4 5 AFIB MAM 6 7 8 9 10 11 12 13 1 o a e...

Page 12: ...150 85 120 98 c 2 AFIB AFIB 65 40 AFIB Microlife MAM AFIB Microlife 3 AT 1 M 2 Time 5 2 M Time 3 WHO 2003 100 60 1 100 120 60 80 2 120 130 80 85 3 130 140 85 90 4 140 160 90 100 5 160 180 100 110 6 18...

Page 13: ...e Easy Microlife Microlife 6 7 4 MAM MAM MAM AK 1 MAM 3 MAM MAM AFIB MAM 1 MAM BK 1 2 3 15 15 Blood Pressure Monitoring 2001 6 145 147 4 1 2 3 4 5 2 3 6 1 7 8 9 BT S 17 22 6 75 8 75 M 22 32 8 75 12 5...

Page 14: ...12 10 AM AN AO 11 12 1 y 1 M AS M 2 30 40 5 AFIB MAM AP AFIB AFIB 6 Microlife Blood Pressure Analyser BPA CD BPA www microlife ru USB B 5 1 CD ROM SETUP EXE 2 3 3 3...

Page 15: ...13 BP A110 RU help 7 M 2 M AS M 17 17 M M 200 M CL M CL 8 AR AR 1 AT 2 3 3 4 1 5 AA oc a NiMH 9 Microlife 6 600 Microlife Microlife 230 1 8 2...

Page 16: ...14 10 ERR 3 1 11 1 ERR 1 ERR 2 ERR 3 ERR 5 ERR 6 MAM MAM HI 300 200 5 LO 40...

Page 17: ...Microlife 13 93 42 EEC 2015 2623 02 06 2015 2 2 5 c 10 10 40 C 15 95 20 55 C 15 95 560 152 x 92 x 42 I V 20 280 40 200 0 299 1 3 5 4 x 1 5 AA 6 600 920 AA 4 IP20 EN 1060 1 3 4 IEC 60601 1 IEC 60601 1...

Page 18: ...16 27 495 325 45 63 8 800 200 33 22...

Page 19: ...17 BP A110 RU...

Page 20: ...iakidis M Diagnostic accuracy of a home blood pressure monitor to detect atrial fibrillation J Hum Hyperten 2009 1 5 2 Wiesel J Fitzig L Herschman Y Messineo FC Detection of Atrial Fibrillation Using...

Page 21: ...19 BP A110 UA 1 2 AFIB AFIB i Microlife AFIB i MAM 3 MAM MAM 4 5 i MAM 6 7 8 9 10 11 12 13 1 a 15...

Page 22: ...B AFIB atria i 65 40 Microlife AFIB i MAM i Microlife AFIB i i i 3 AT 1 M 2 Time 5 2 M Time 3 4 Time 5 Time 3 Microlife 2003 100 60 1 100 120 60 80 2 120 130 80 85 3 130 140 85 90 4 140 160 90 100 5 1...

Page 23: ...ife 6 7 4 MAM MAM MAM AK 1 MAM 3 MAM MAM i MAM 1 MAM BK 1 2 3 15 15 Blood Pressure Monitoring 2001 6 145 147 4 1 2 3 4 5 2 3 6 1 7 8 9 BT 10 AM AN AO 11 12 1 S 17 22 6 75 8 75 M 22 32 8 75 12 5 L 32 4...

Page 24: ...22 y 1 M AS M 2 30 40 5 i MAM AP i 6 Microlife Blood Pressure Monitor BPA Microlife BPA D www microlife com ua USB B 5 1 CD D ROM SETUP EXE 2 H 3 3 i i help 7 M 2 M AS M 17 17 i...

Page 25: ...23 BP A110 UA M M a 200 i y i M CL M CL 8 AR AR 1 AT 2 3 3 4 1 5 AA NiMH i i i e 9 Microlife DC 6 600 Microlife Microlife 230 1 8 2 10 ERR 3 ERR 1...

Page 26: ...24 1 11 1 2 ERR 2 ERR 3 ERR 5 ERR 6 MAM MAM HI 300 200 5 LO 40...

Page 27: ...icrolife 12 5 o Microlife 13 93 42 EEC 10 40 C 15 95 20 55 C 15 95 560 152 x 92 x 42 I V 20 280 40 200 0 299 1 3 5 4 x 1 5 AA 6 600 920 AA 4 IP IP20 EN 1060 1 3 4 IEC 60601 1 IEC 60601 1 2 EMC IEC 606...

Reviews: