select standard mode, slide the MAM switch 

AK

 on the side of the 

device downwards to position «1» and to select MAM mode, slide 

this switch upwards to position «3».

MAM mode

In MAM mode, 3 measurements are automatically taken in succes-

sion and the result is then automatically analysed and displayed. 

Because blood pressure constantly fluctuates, a result determined 

in this way is more reliable than one produced by a single 

measurement. AFIB detection is only activated in MAM mode.

After pressing the ON/OFF button 

1

, the MAM-symbol 

BK

 

appears in the display.

The bottom, right hand section of the display shows a 1, 2 or 3 to 

indicate which of the 3 measurements is currently being taken.

There is a break of 15 seconds between the measurements 

(15 seconds are adequate according to «Blood Pressure Moni-

toring, 2001, 6:145-147» for oscillometric instruments). A count 

down indicates the remaining time.

The individual results are not displayed. Your blood pressure 

will only be displayed after all 3 measurements are taken.

Do not remove the cuff between measurements.

If one of the individual measurements was questionable, a 

fourth one is automatically taken.

4. Taking a Blood Pressure Measurement using this 

Device

Checklist for taking a reliable measurement

1. Avoid activity, eating or smoking immediately before the 

measurement.

2. Sit down for at least 5 minutes before the measurement and relax.

3. Always measure on the same arm (normally left).

4. Remove close-fitting garments from the upper arm. To avoid 

constriction, shirt sleeves should not be rolled up - they do not 

interfere with the cuff if they are laid flat.

5. Always ensure that the correct cuff size is used (marking on the 

cuff).

Fit the cuff closely, but not too tight.

Make sure that the cuff is positioned 2 cm above the elbow.

The 

artery mark

 located on the cuff (ca. 3 cm long bar) must 

lie over the artery which runs down the inner side of the arm.

Support your arm so it is relaxed.

Ensure that the cuff is at the same height as your heart.

6. Press the ON/OFF button 

1

 to start the measurement.

7. The cuff will now pump up automatically. Relax, do not move 

and do not tense your arm muscles until the measurement 

result is displayed. Breathe normally and do not talk.

8. When the correct pressure is reached, the pumping stops and the 

pressure falls gradually. If the required pressure was not reached, 

the device will automatically pump some more air into the cuff.

9. During the measurement, the heart symbol 

BT

 flashes in the 

display every time a heartbeat is detected.

10.The result, comprising the systolic 

AM

 and the diastolic 

AN

 

blood pressure and the pulse rate 

AO

 is displayed. Note also the 

explanations on further display symbols in this booklet.

11.When the device has finished measuring, remove the cuff.

12.Enter the result in the enclosed blood pressure pass and switch 

off the device. (The monitor does switch off automatically after 

approx. 1 min.).

How not to store a reading

As soon as the reading is displayed press and hold the ON/OFF 

button 

1

 until «

M

» 

AS

 is flashing. Confirm to delete the reading by 

pressing the M-button 

2

.

You can stop the measurement at any time by pressing the 

ON/OFF button (e.g. if you feel uneasy or an unpleasant 

pressure sensation). 

If the systolic blood pressure is known to be very high

it can be an advantage to set the pressure individually. Press 

the ON/OFF button after the monitor has been pumped up to 

a level of approx. 30 mmHg (shown on the display). Keep the 

button pressed until the pressure is about 40 mmHg above 

the expected systolic value – then release the button. 

5. Appearance of the Atrial Fibrillation Indicator for 

early Detection (only in MAM mode)

This symbol 

AP

 indicates that an atrial fibrillation was detected 

during the measurement. This device is able to detect atrial fibril-

lation (AFIB). If AFIB is present during blood pressure measure-

ment, the AFIB Indicator is displayed. In this case, the result may 

deviate from your normal blood pressure. It is highly recom-

mended to take an additional measurement an hour later to 

increase the specificity of the detection. In most cases, this is no 

cause for concern. However, if the symbol appears on a regular 

basis (e.g. several times a week with measurements taken daily) 

we advise you to visit your doctor. Please, provide the following 

explanation:

Summary of Contents for BP A110

Page 1: ...ww microlife com Asia Microlife Corporation 9F 431 RuiGang Road NeiHu Taipei 11492 Taiwan R O C Tel 886 2 8797 1288 Fax 886 2 8797 1283 Email service microlife com tw www microlife com North Central S...

Page 2: ...5 BP A110 UA o Microlife 12 5 o Microlife 13 93 42 EEC 10 40 C 15 95 20 55 C 15 95 560 152 x 92 x 42 I V 20 280 40 200 0 299 1 3 5 4 x 1 5 AA 6 600 920 AA 4 IP IP20 EN 1060 1 3 4 IEC 60601 1 IEC 60601...

Page 3: ...etects patients with AFIB at 97 100 certainty 1 2 Please read through these instructions carefully so that you under stand all functions and safety information We want you to be happy with your Microl...

Page 4: ...h blood pressure values can damage your health and must be treated by your doctor Always discuss your values with your doctor and tell him her if you have noticed anything unusual or feel unsure Never...

Page 5: ...tery compartment AT Setting the date and time 1 After the new batteries are fitted the year number flashes in the display You can set the year by pressing the M button 2 To confirm and then set the mo...

Page 6: ...7 The cuff will now pump up automatically Relax do not move and do not tense your arm muscles until the measurement result is displayed Breathe normally and do not talk 8 When the correct pressure is...

Page 7: ...y capacity of 200 is not exceeded When the memory is full the old values are automatically overwritten with new ones Values should be evaluated by a doctor before the memory capacity is reached otherw...

Page 8: ...fety and protection This device may only be used for the purposes described in these instructions The manufacturer cannot be held liable for damage caused by incorrect application This device comprise...

Page 9: ...ase The guarantee is valid only on presentation of the guar antee card completed by the dealer see back confirming date of purchase or the receipt Batteries cuff and parts that become worn with use ar...

Page 10: ...riakidis M Diagnostic accuracy of a home blood pressure monitor to detect atrial fibrillation J Hum Hyperten 2009 1 5 2 Wiesel J Fitzig L Herschman Y Messineo FC Detection of Atrial Fibrillation Using...

Page 11: ...9 BP A110 RU 1 2 AFIB AFIB AFIB Microlife MAM 3 MAM MAM 4 5 AFIB MAM 6 7 8 9 10 11 12 13 1 o a e...

Page 12: ...150 85 120 98 c 2 AFIB AFIB 65 40 AFIB Microlife MAM AFIB Microlife 3 AT 1 M 2 Time 5 2 M Time 3 WHO 2003 100 60 1 100 120 60 80 2 120 130 80 85 3 130 140 85 90 4 140 160 90 100 5 160 180 100 110 6 18...

Page 13: ...e Easy Microlife Microlife 6 7 4 MAM MAM MAM AK 1 MAM 3 MAM MAM AFIB MAM 1 MAM BK 1 2 3 15 15 Blood Pressure Monitoring 2001 6 145 147 4 1 2 3 4 5 2 3 6 1 7 8 9 BT S 17 22 6 75 8 75 M 22 32 8 75 12 5...

Page 14: ...12 10 AM AN AO 11 12 1 y 1 M AS M 2 30 40 5 AFIB MAM AP AFIB AFIB 6 Microlife Blood Pressure Analyser BPA CD BPA www microlife ru USB B 5 1 CD ROM SETUP EXE 2 3 3 3...

Page 15: ...13 BP A110 RU help 7 M 2 M AS M 17 17 M M 200 M CL M CL 8 AR AR 1 AT 2 3 3 4 1 5 AA oc a NiMH 9 Microlife 6 600 Microlife Microlife 230 1 8 2...

Page 16: ...14 10 ERR 3 1 11 1 ERR 1 ERR 2 ERR 3 ERR 5 ERR 6 MAM MAM HI 300 200 5 LO 40...

Page 17: ...Microlife 13 93 42 EEC 2015 2623 02 06 2015 2 2 5 c 10 10 40 C 15 95 20 55 C 15 95 560 152 x 92 x 42 I V 20 280 40 200 0 299 1 3 5 4 x 1 5 AA 6 600 920 AA 4 IP20 EN 1060 1 3 4 IEC 60601 1 IEC 60601 1...

Page 18: ...16 27 495 325 45 63 8 800 200 33 22...

Page 19: ...17 BP A110 RU...

Page 20: ...iakidis M Diagnostic accuracy of a home blood pressure monitor to detect atrial fibrillation J Hum Hyperten 2009 1 5 2 Wiesel J Fitzig L Herschman Y Messineo FC Detection of Atrial Fibrillation Using...

Page 21: ...19 BP A110 UA 1 2 AFIB AFIB i Microlife AFIB i MAM 3 MAM MAM 4 5 i MAM 6 7 8 9 10 11 12 13 1 a 15...

Page 22: ...B AFIB atria i 65 40 Microlife AFIB i MAM i Microlife AFIB i i i 3 AT 1 M 2 Time 5 2 M Time 3 4 Time 5 Time 3 Microlife 2003 100 60 1 100 120 60 80 2 120 130 80 85 3 130 140 85 90 4 140 160 90 100 5 1...

Page 23: ...ife 6 7 4 MAM MAM MAM AK 1 MAM 3 MAM MAM i MAM 1 MAM BK 1 2 3 15 15 Blood Pressure Monitoring 2001 6 145 147 4 1 2 3 4 5 2 3 6 1 7 8 9 BT 10 AM AN AO 11 12 1 S 17 22 6 75 8 75 M 22 32 8 75 12 5 L 32 4...

Page 24: ...22 y 1 M AS M 2 30 40 5 i MAM AP i 6 Microlife Blood Pressure Monitor BPA Microlife BPA D www microlife com ua USB B 5 1 CD D ROM SETUP EXE 2 H 3 3 i i help 7 M 2 M AS M 17 17 i...

Page 25: ...23 BP A110 UA M M a 200 i y i M CL M CL 8 AR AR 1 AT 2 3 3 4 1 5 AA NiMH i i i e 9 Microlife DC 6 600 Microlife Microlife 230 1 8 2 10 ERR 3 ERR 1...

Page 26: ...24 1 11 1 2 ERR 2 ERR 3 ERR 5 ERR 6 MAM MAM HI 300 200 5 LO 40...

Page 27: ...icrolife 12 5 o Microlife 13 93 42 EEC 10 40 C 15 95 20 55 C 15 95 560 152 x 92 x 42 I V 20 280 40 200 0 299 1 3 5 4 x 1 5 AA 6 600 920 AA 4 IP IP20 EN 1060 1 3 4 IEC 60601 1 IEC 60601 1 2 EMC IEC 606...

Reviews: