How
to
Set
Up
and
Run
the
FG5
3
3.1.2.2.WEO Model 100 Laser
•
Turn on the power (main power and HV key switches).
•
Select the proper iodine peak.
•
Set the servo control to AUTO.
•
Do not adjust any other controls.
Nominal control settings are:
Signal Monitor Section
:
Meter Select:
1F
Gain
1
Time Constant
1
Temperature Control Section:
Body Temp Mode
OFF
Temperature Control Section
Meter
BIAS
Bias Voltage
Set to 0V (meter)
•
Allow the laser to warm up for at least two hours (or until the
temperature of the
laser has stabilized) before beginning observations.
3.1.3.DROPPING CHAMBER TRIPOD
14. Remove the tripod tray from the Superspring case and place it carefully
upside down on the floor.
15. Remove the three tripod legs from the dropping chamber case and attach
them to the tray.
16. Turn the tripod upright and tighten the legs by turning the large 3-lobe
knobs clockwise.
At this point the interferometer will be used to support the dropping
chamber tripod.
3-4
Summary of Contents for FG5
Page 22: ...Design Components and Function 2 Figure 2 10 The Superspring 2 14 ...
Page 31: ...Design Components and Function 2 Figure 2 13 Rotation Monitor 2 23 ...
Page 32: ......
Page 42: ...How to Set Up and Run the FG5 3 3 10 ...
Page 44: ...How to Set Up and Run the FG5 3 Figure 3 2 V Post 3 12 ...
Page 53: ...How to Set Up and Run the FG5 3 1 Backup the data 2 Shut off computer power 3 21 ...
Page 87: ...Adjustment and Maintenance 4 4 29 ...
Page 91: ...Adjustment and Maintenance 4 4 33 ...
Page 104: ...Troubleshooting 5 5 2 ...
Page 117: ...Troubleshooting 5 5 15 ...
Page 131: ...Checklists and Logs Appendix D 9 3 ...
Page 140: ...Checklists and Logs Appendix D Table 9 6 Replacing Drive Belt 9 12 ...
Page 145: ...Checklists and Logs Appendix D Table 9 10 Replace Linear Bearings 9 17 ...
Page 149: ...Checklists and Logs Appendix D Table 9 13 Replace Shaft Encoder 9 21 ...