Checklists and Logs
Appendix D
9.4.4.Adjust Belt Tension
Adjusting Belt Tension Checklist
Tighten belt adjustment screw until belt is straight.
Tension belt by tightening adjustment screw three turns (until
tension spring is just short of coil bind). or tension belt using torque
wrench.
Table 9-7 Adjusting Belt Tension
9.4.5.Replace Ferrofluidic Vacuum
Feedthrough
Replacing Ferrofluidic Vacuum Feedthrough Checklist
Remove dropping chamber cover.
Loosen 4-40 screws on Helical coupling (between motor and
feedthrough).
Remove ¼-28 screws which attach motor to Conflat flange.
Remove motor mount assembly (motor, Helical coupling, travel
lock plate, and shaft encoder) from Conflat flange.
Loosen screw on Helical coupling (where it attaches to drive pulley
shaft).
9-13
Summary of Contents for FG5
Page 22: ...Design Components and Function 2 Figure 2 10 The Superspring 2 14 ...
Page 31: ...Design Components and Function 2 Figure 2 13 Rotation Monitor 2 23 ...
Page 32: ......
Page 42: ...How to Set Up and Run the FG5 3 3 10 ...
Page 44: ...How to Set Up and Run the FG5 3 Figure 3 2 V Post 3 12 ...
Page 53: ...How to Set Up and Run the FG5 3 1 Backup the data 2 Shut off computer power 3 21 ...
Page 87: ...Adjustment and Maintenance 4 4 29 ...
Page 91: ...Adjustment and Maintenance 4 4 33 ...
Page 104: ...Troubleshooting 5 5 2 ...
Page 117: ...Troubleshooting 5 5 15 ...
Page 131: ...Checklists and Logs Appendix D 9 3 ...
Page 140: ...Checklists and Logs Appendix D Table 9 6 Replacing Drive Belt 9 12 ...
Page 145: ...Checklists and Logs Appendix D Table 9 10 Replace Linear Bearings 9 17 ...
Page 149: ...Checklists and Logs Appendix D Table 9 13 Replace Shaft Encoder 9 21 ...