Checklists and Logs
Appendix D
Connect Superspring controller to Superspring.
Drive zero-positioning motor (ZPM) until sphere voltage is zero.
Measure gap between two levers near ZPM.
Adjust coarse adjustment screw so motor will be in center of travel
position (6 ± 0.5 mm gap between lever arms).
Make sure sphere voltage goes +/- to assure true center position.
Carefully align beam holes in top lever with holes in test mass.
Recheck centering of test mass in cage and adjust bubble levels.
Table 9-20 Assembling Superspring
9.6.7.Replacing Focus Lever Motor
9-30
Replacing Focus Lever Motor Checklist
Remove nut on focus adjustment screw.
Lower focus lever motor assembly off fulcrum lever.
Loosen set screw holding hex bushing to motor shaft.
Raise bushing out of the way.
Summary of Contents for FG5
Page 22: ...Design Components and Function 2 Figure 2 10 The Superspring 2 14 ...
Page 31: ...Design Components and Function 2 Figure 2 13 Rotation Monitor 2 23 ...
Page 32: ......
Page 42: ...How to Set Up and Run the FG5 3 3 10 ...
Page 44: ...How to Set Up and Run the FG5 3 Figure 3 2 V Post 3 12 ...
Page 53: ...How to Set Up and Run the FG5 3 1 Backup the data 2 Shut off computer power 3 21 ...
Page 87: ...Adjustment and Maintenance 4 4 29 ...
Page 91: ...Adjustment and Maintenance 4 4 33 ...
Page 104: ...Troubleshooting 5 5 2 ...
Page 117: ...Troubleshooting 5 5 15 ...
Page 131: ...Checklists and Logs Appendix D 9 3 ...
Page 140: ...Checklists and Logs Appendix D Table 9 6 Replacing Drive Belt 9 12 ...
Page 145: ...Checklists and Logs Appendix D Table 9 10 Replace Linear Bearings 9 17 ...
Page 149: ...Checklists and Logs Appendix D Table 9 13 Replace Shaft Encoder 9 21 ...