How
to
Set
Up
and
Run
the
FG5
3
3.1.2.1.Model AL-1 Laser:
3.1.2.1.1.Warm Up
•
Set the LOOP switch to the OPEN position.
•
Turn the switch on the laser controller to MANUAL.
•
DO NOT ADJUST THE COARSE GAIN KNOB.
•
Turn on the main power switch and the key switch for the laser tube
HV. Both lights on the laser controller panel should come on.
•
Set the heater current to 0.3 using the front panel monitor to view
current..
•
Allow the laser to warm up for at least half an hour before locking.
3.1.2.1.2.Operation
•
After the laser is warm, the LOOP switch should be closed.
•
The control switch should be set to REMOTE so that the computer can
control the red/blue side-lock choice.
3-3