DMC 2000S Dosimeter
15-00007 Page 62 of 65
12.4 Mechanical characteristics
!
dimensions:
#
Width: 44 mm; 1.76 in.
#
Height: 83 mm; 3.26 in.
#
depth: 9 mm .35 in. (18 mm; .70 in. at the display)
!
Weight: approximately 50 g; 1.76 oz.
!
drop resistant (up to 25 drops) from a height of 1.5 m (4’9”) onto a hard surface
12.5 Environmental characteristics
!
operational from - 10 to + 50 °C with a lower than
±
20 % variation in the dose
equivalent measurement
!
storage temperature - 31°C to + 71°C (without battery)
12.6 Link to LDM 2000 reader
!
Pass-by data exchange for activation, deactivation, in-motion data acquisition, reading
and writing of parameters
#
normal range of 1.2 m (3’9”) for pass-by readings (2.4 m (7’8”) if equipped with
external antenna)
#
range < 0.5 m for configuration (1’6”)
12.7 Storage
In the event the dosimeter will not be used for a prolonged period of time, putting the
dosimeter into “storage” reduces its power consumption to a minimum (
dosimeter
storage
mode).
In this mode, messages are no longer displayed.
The
LDM 2000
reader is required to put the dosimeter in the storage mode.