DMC 2000S Dosimeter
15-00007 Page 42 of 65
8.3.1.2. Dose alarm threshold
modifiable in both the satellite (with an LDM-101 or LDM 2000 and the appropriate
software) and autonomous modes
This parameter corresponds to the dose equivalent value above which an alarm will be
triggered.
This value can be set from 1.0 * 10-1 to 9.99 * 10+5 mRem.
8.3.1.3. Rate alarm threshold
modifiable in both the satellite (with an LDM-101 or LDM 2000 and the appropriate
software) and autonomous modes
This parameter corresponds to the dose rate value above which an alarm will be
triggered.
This value can be set from 1.0 * 10-1 to 9.99 * 10+5 mRem.
8.3.1.4. Time alarm threshold
modifiable only in the satellite mode, with an LDM-101 or LDM 2000 and the
appropriate software
This parameter corresponds to the time period operating in the
measurement
mode after
which an alarm will be triggered.
This value can be set from 0-hr. 01 min. to 99-hr. 59 min.
If this value is set to
00 hr. 00
, the
time
alarm will not be triggered.