DMC 2000S Dosimeter
15-00007 Page 47 of 65
8.4.2.3.Step “ I ”
The flashing “I” displayed to the left of the “functional indicators” identifies this step. It
allows modification of the selected digit of the current parameter by
increments
.
!
pressing and releasing changes the value by 1 increment
!
pressing and holding the selector button takes you to the next step
Note
: for “general” parameters, pressing and releasing the selector button allows
selection of the available values (for example: “RESET “ or “TOTAL”)
Steps
“
I ”
and
“d”
are displayed after step
“ P ”
, but not in any particular order.
The DMC 2000S dosimeter automatically switches back to the
pause
mode after 10
seconds of inactivity on the selector button.
8.4.3.
Example for setting parameters
!
access the
modification
mode as described in the section
, page 44.
!
go to the parameter you want to modify
:
#
by successively pressing and releasing the selector button
example shown: dose alarm threshold
the parameter is displayed and is preceded by a flashing
“
P
”
(parameter setting mode is
active)
!
to access
the digit you want to change:
#
press and hold the selector button until the flashing
“
d
”
is displayed (the digit which
can be changed is flashing)
#
press and release the selector button until the digit you want to change is flashing
in the example shown, the digit to be modified flashes as soon as the
“d”
is displayed:
!
to change
the
value of the selected digit