DMC 2000S Dosimeter
15-00007 Page 43 of 65
8.3.1.5. Dose pre-alarm warning threshold
modifiable in both the satellite (with an LDM-101 or LDM 2000 and the appropriate
software) and autonomous modes
This parameter corresponds to the dose equivalent value above which a pre-alarm will be
triggered.
This value can be set from 1.0 * 10-1 to 9.99 * 10+5 mRem; it should always be lower
than the dose equivalent alarm threshold.
If this value is set to
0
(
0.00.10
0
), the
dose
pre-alarm will not be triggered.
8.3.1.6. Rate pre-alarm warning threshold
modifiable in both the satellite (with an LDM-101 or LDM 2000 and the appropriate
software) and autonomous modes
This parameter corresponds to the dose rate value above which a pre-alarm will be
triggered.
This value can be set from 1.0 * 10-1 to 9.99 * 10+5 mRem; it should always be lower
than the dose rate alarm threshold.
If this value is set to
0
(
0.00.10
0
), the
rate
pre-alarm will not be triggered
.