Page 95 P/N 03521015 Rev U1
Bio-Medical Devices Intl
17171 Daimler Street Irvine, CA 92614 800-443-3842 www.maxair-systems.com
WARNING
Use only if package is received unopened and contents are
undamaged. If damage is noted, contact the shipper for replace-
ment or repair.
Prior to using any MAXAIR
®
System or component, be sure to be
familiar with the system’s NIOSH approved configuration.
DO NOT use if any component is damaged. If any components are
damaged or contaminated and therefore unfit for safe and effective
use, they should be replaced immediately.
Only trained and experienced personnel who have read and under-
stand the User’s Instructions should use this product.
Failure to follow the User’s Instructions, these and in the MAXAIR
Systems User’s Instructions, P/N 03521015, may be hazardous to
the user’s health.
The institution using this product in any application is responsible
for determining the appropriateness of this equipment relative to
regulatory requirements. Bio-Medical Devices Intl, Inc. does not
recommend the appropriate systems for a particular institution or
facility.
Use only MAXAIR Systems/ NIOSH approved compatible
components.
NOT for use in atmospheres immediately dangerous to life or health
(IDLH), and atmospheres containing less than 19.5% oxygen, or
more than 25% oxygen.
If you need more information, contact your BMDI Sales
Representative, or call BMDI customer service at 1-800-443-3842.
Symbol Definitions
Warning, Caution, or Note
Order Number Part Number
O.N.
P/N
Intended Use
The MAXAIR HFR FCC (High Fluid Resistant Filter Cover Cap) is
to provide high fluid resistance protection for MAXAIR Hood Filters
against airborne and contact transmissible contaminants.
Specifications
Temperature Limits: Melting Point approximately 150-160
o
F
Materials
Polypropylene Co-Polymer
O.N. 2061-05
P/N 01021609
Regulatory
NIOSH
For Use With
The HFR FCC is to be used with MAXAIR XP Hood System
configurations.
3. To fully secure the HFR FCC, pull it down in back with fingers
while pushing up on under side of Helmet back with thumb.
1. Place the front of the HFR FCC
over the front of the Helmet and
continue moving it down towards
the back.
2. Continue bringing the
HFR FCC down over the
Helmet so that it covers
the HLF.
1. Hold HFR FCC front and lift
its rear up over Helmet.
Doff HFR FCC
2. Remove HFR FCC completely
and decon for next use.
Don HFR FCC (on to an XP Hood with HLF)
25. High Fluid Resistance Filter Cover Cap (HFR FCC)