Rad-67
Chapter 4: Operation
www.masimo.com
37
Masimo
delivers enhanced protection against erroneous pulse rate and arterial oxygen
saturation readings when a sensor becomes inadvertently detached from a
patient due to excessive movement.
MAX (Maximum Sensitivity)
MAX is the recommended sensitivity mode for patients with low perfusion or
when a
low perfusion
message displays in APOD or NORM mode. MAX mode is
not recommended for care areas where patients are not monitored visually, such
as medical-surgical floors. It is designed to display data at the measuring site
when the signal may be weak due to decreased perfusion. When a sensor
becomes detached from a patient, it will have compromised protection against
erroneous pulse rate and arterial saturation readings.
Understanding Windows
The following information describes how to customize the information viewed on the
Main
Screen
.
Customizing Windows
Windows can be customized by expanding and minimizing parameters and measurements.
When a parameter is minimized, it is only displayed in the Well with its Numeric Value and
Parameter Label. When a parameter is expanded, it will be shown in the Parameter display.
To expand a parameter or measurement
Order Instruction
Step 1 Touch and hold the Numeric Value until it dims.
Step 2 Drag the Numeric Value over the Parameter Display.
Step 3 Release the Numeric Value.