Rad-67
Chapter 5: Spot-Check
www.masimo.com
59
Masimo
Place the sensor on the patient's digit. Refer to the
Directions for Use for the specific sensor for proper site
selection and application.
After the sensor is placed, the Rad-67 looks for a pulse.
Once a pulse is detected, the spot-check begins.
Note:
If the sensor is already on a digit and a pulse is
detected, this step is skipped and the spot-check begins.
The spot-check requires user input if SpHb is to be
included and measured. Select the
Measure SpHb
button
to include SpHb in the spot-check.
After the sensor initializes, a pleth is displayed and spot-
check measurement begins to display available values.
Note:
In the following situations, the ability to include
SpHb is not available and the
Measure SpHb
button is
grayed out.
When
skip
is selected instead of entering a patient
label (when required), age and gender.
When the entered patient age is 17 or less. SpHb
measurement can only be performed on patients 18
years of age or older.