Chapter 7: Troubleshooting
www.masimo.com
73
Masimo
Symptom
Potential Causes
Next Steps
Device does not detect
that the sensor is
connected
Sensor not properly
connected to device.
Improper placement
of sensor.
Damaged sensor.
Internal failure.
Remove and reconnect sensor.
Ensure the connector is fully
connected to the device.
Reapply sensor to the patient.
Refer to sensor
Directions For
Use
.
Replace sensor.
Turn Rad-67 Off and On. See
Powering the Rad-67 ON and
OFF
on page 28.
Contact Masimo service. See
Contacting Masimo
on page
Device does not
communicate to other
external devices
through wireless
connection
External device is
not compatible.
Wi-Fi is not turned
on and/or not
correctly configured.
Location does not
have wireless
availability.
Connected network
is not available.
Internal failure.
Check external device
compatibility.
Check that the wireless feature
is on and correctly configured.
See
Wi-Fi
on page 49.
Check wireless availability for
location.
Check network settings and
availability.
Contact Masimo service. See
Contacting Masimo
on page