Chapter 7: Troubleshooting
www.masimo.com
72
Masimo
Symptom
Potential Causes
Next Steps
Device screen is blank
The device is in low
power mode to
conserve power.
The device is Off.
The brightness
display is not
correct.
Battery may be
depleted.
Internal failure.
Touch the screen to turn the
display On.
Turn Rad-67 Off and On. See
Powering the Rad-67 ON and
OFF
on page 28.
Adjust the brightness setting.
See
Brightness
on page 52.
Check AC power connection
and recharge.
Contact Masimo service. See
Contacting Masimo
on page
Touchscreen/Buttons
do not respond when
pressed
EMI (Electro
Magnetic
Interference).
Internal failure.
Turn Rad-67 Off and On. See
Powering the Rad-67 ON and
OFF
on page 28.
Relocate the device from other
devices that may cause
electromagnetic interference.
Contact Masimo service. See
Contacting Masimo
on page
Battery run time
significantly reduced
Battery not fully
charged.
Battery damaged.
Battery capacity
effected.
Check battery charge level
indicator.
Check battery is fully charged.
Contact Masimo service. See
Contacting Masimo
on page
Device does not detect
that patient cable is
connected
Cable connector not
properly connected
to the device.
Damaged connector.
Damaged cable.
Cable expired.
Internal failure.
Remove and reconnect cable.
Ensure the connector is fully
connected to the device.
Replace cable.
Contact Masimo service. See
Contacting Masimo
on page