Chapter 8: Specifications
www.masimo.com
84
Masimo
Guidance and Manufacturer's Declaration-Electromagnetic
Immunity
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The ME Equipment is intended for use in the electromagnetic environment specified
below. The customer or the user of the ME Equipment should assure that it is used in such
an environment.
Immunity Test
IEC
60601
Test Level
Compliance
Level
Electromagnetic Environment -
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+6 kV
contact
+6 kV contact Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
+8 kV air
+8 kV air
Electrical fast
transient/ burst
IEC 61000-4-4
+/- 2 kV
for power
lines
+/- 2 kV for
power lines
Mains power quality should be that of
a typical commercial or hospital
environment.
+/- 1 kV
for input/
output
lines
+/- 1 kV for
input/ output
lines
Surge
IEC 61000-4-5
+/-1 kV
line(s) to
line(s)
+/-1 kV line(s)
to line(s)
Mains power quality should be that of
a typical commercial or hospital
environment.
+/- 2 kV
line(s) to
earth
+/- 2 kV
line(s) to
earth
Voltage dips, short
interruptions, and
voltage variations
on power supply
input lines
IEC 61000-4-11
100% dip
in mains
voltage
for 0.5
cycle
100% dip in
mains voltage
for 0.5 cycle
Mains power quality should be that of
a typical commercial or hospital
environment.
60% dip
in mains
voltage
for 5 cycle
60% dip in
mains voltage
for 5 cycle