6.10 RTD Measurement and Simulation
1)
Pt10 …… Pt1000, -200 …… 850 ºC
Range
Resolution
Accuracy
-
200….200 °C
Pt10…Pt400:
0.01
°C
Pt500,Pt1000:
0.1
°C
4 wire Measurement: 0.15°C
Simulation*: 0.15 °C
200….600 °C
4 wire Measurement: 0.2 °C
Simulation*: 0.25 °C
600….850 °C
4 wire Measurement: 0.3 °C
Simulation*: 0.35 °C
2) Ni100, -
60 …… 180 ºC
Range
Resolution
Accuracy
-
60….180 °C
0.01 °C
4 wire Measurement: 0.1 °C
Simulation*: 0.15 °C
3) Ni120, -
80 …… 260 ºC
Range
Resolution
Accuracy
-
80….260 °C
0.01 °C
4 wire Measurement: 0.1 °C
Simulation*: 0.15 °C
4) Cu10, -
200 …… 260 ºC
Range
Resolution
Accuracy
-
200….260 °C 0.01 °C
4 wire Measurement: 0.2 °C
Simulation*: 0.8 °C
Note:
* Specification valid with an excitation current >0.2 mA (0...400 ohm),
>0.1 mA (400...4000 ohm)
** Read accuracy is based on 4-wire input. For 3-wire RTD
measurements, assuming all three
RTD leads are matched, add 1.0 °C (Pt10 and Cu10), 0.6 °C (Pt50
and Cu50), and 0.4 °C
(Other RTD types) to the specifications.
6.11 Available RTD Types
1) Pt10 (385)
6) Pt500 (385)
11) Ni120 (672)
2) Pt50 (385)
7) Pt1000 (385)
12) Cu10 (427)
3) Pt100 (385)
8) Pt100 (3926)
13) Cu50 (427)
4) Pt200 (385)
9) Ni100 (672)
14) Cu100 (427)
5) Pt400 (385)
10) Ni100 (618)