Parameter Name Description / Options
I/P Type
(Input Type)
Measure Input Type
Available Options:
mA : 0.000 to 24.000 mA DC
mA(24V) : 0.000 to 24.000 mA DC
V : 0.000 to 30.000 V DC
Pulse : 0 to 999999 Pulses
Frequency : 0.0143 to 50000Hz
Input Range
Low (0%)
Low Range for Measure Input. Only for mA, mA(24V) and V
Range:
Default Input Low to
Input Range High (100%)
This parameter is enabled, if
Main Display
in
MENU
→
DISPLAY
→
EM SETUP
is set to
Percentage
or
Scaled.
Input Range
High (100%)
High Range for Measure Input. Only for mA, mA(24V) and V
Range:
Input Range Low(0%)
to Default Input High
This parameter is enabled, if
Main Display
in
MENU
→
DISPLAY
→
EM SETUP
is set to
Percentage
or
Scaled.
Scaled Input Range
Low (0%)
Scaling Low Range for Measure Input. Only for mA, mA(24V) and V
Range:
-99999 to
Scaled Input Range High (100%)
Decimal Point for this EditBox can be changeable.
This parameter is enabled, if
Main Display
in
MENU
→
DISPLAY
→
EM SETUP
is set to
Scaled.
Scaled Input Range
High (100%)
Scaling High Range for Measure Input. Only for mA, mA(24V) and V
Range:
Scaled Input Range Low(0%)
to 99999
Decimal Point for this EditBox can be changeable.
This parameter is enabled, if
Main Display
in
MENU
→
DISPLAY
→
EM SETUP
is set to
Scaled.
TF
(Transfer Function)
Transfer Function for Scaling. Only for mA, mA(24V) and V
Available Options:
Linear
x^2 (x
2
)
x^(1/2) (
√x)
This parameter is enabled, if
Main Display
in
MENU
→
DISPLAY
→
EM SETUP
is set to
Scaled.
Trigger Edge
Trigger Edge for Pulse Measure input