36
SP2000, SP2300, SP2400 | 1.00.03
www.mc-techgroup.com
16.4.1
OPTION DRIVE UNIT 234B FOR THE SOLENOID VALVE UNITS
The drive unit
234B
is used for control of the solenoid valves in the solenoid valve units for a clocked back purging.
The unit is mounted on a cap rail inside the FRP protection housing of the solenoid valve unit and is delivered
electrically connected.
16.4.1.1
Function and adjustment of the clock generator K3
The control of the back purge action is made via the electronic clock generator
K3
(Typ CT-
MXS.22) that is equipped with 2 separate time adjustment possibilities for the impulse and
the pause duration. During the impulse duration, the probe is pack purged.
In the front of the clock generator are situated the adjustment possibilities for the time ranges
(Range) and the potentiometer for the fine adjustment (Time). Thus, any back purge and
pause durations can be separately adjusted. For the fine adjustment (Time) the scale is valid
which corresponds to the colour of the time range adjustment (Range).
Example:
The back purging shall be effected every 12 hours for totally 10 seconds.
Duration of the back purging: 10 s
Pause: 12 h
Adjustment „Duration of the back purging“ (10 s): Set the potentiometer (Time 2) to 10 (white
scale) and the time range (Range 2) to 10 s.
Adjustment „Pause“ (12 h): Set the time range (Range 1) to 30 h and the potentiometer (Time
1) to 12 (orange scale).
16.4.1.2
Function and adjustment of the clock generator K2
Via the clock generator
K2
(EZ12TI), the solenoid valve
Y2
(figure 15 and 16) for the feeding of back purge gas is
controlled. Here, you can adjust the lengths of the back purge impulses and the pause durations between the pack
purge impulses.
In the front of the clock generator
K2
the switch for the time basis T and the multiplicator xT1
and xT2 is situated. The duration of the back purging, eg. 1 s is set with xT1and the duration of
the pause between the back purge blows is set with xT2, eg 1 s.
Example:
Clock time and pause duration: 1 s
Adjust the time basis T at 0.5 s and the multiplicator xT1 and xT2 at 2.
(2 x 0.5 s = 1 s)
Summary of Contents for SP Series
Page 48: ...48 SP2000 SP2300 SP2400 1 00 03 www mc techgroup com Figure 24 SP2000 H basic version ...
Page 49: ...www mc techgroup com SP2000 SP2300 SP2400 1 00 03 49 Figure 25 SP2000 H with options ...
Page 50: ...50 SP2000 SP2300 SP2400 1 00 03 www mc techgroup com Figure 26 SP2300 H ...
Page 51: ...www mc techgroup com SP2000 SP2300 SP2400 1 00 03 51 Figure 27 SP2300 H filter elements ...
Page 53: ...www mc techgroup com SP2000 SP2300 SP2400 1 00 03 53 Figure 29 Heated sample tubes SP30 H ...
Page 54: ...54 SP2000 SP2300 SP2400 1 00 03 www mc techgroup com ...
Page 56: ...56 SP2000 SP2300 SP2400 1 00 03 www mc techgroup com Figure 31 SP2000 H 3VA MS NC B ...
Page 57: ...www mc techgroup com SP2000 SP2300 SP2400 1 00 03 57 Figure 32 SP2000 H 3VA MS NC C ...
Page 58: ...58 SP2000 SP2300 SP2400 1 00 03 www mc techgroup com Figure 33 SP2000 H 3VA MS NO B ...
Page 59: ...www mc techgroup com SP2000 SP2300 SP2400 1 00 03 59 Figure 34 SP2000 H 3VA MS NO C ...