11.2
TEMPERATURE CONTROLLER
In standard version -
H
, or with option
–H320/C
temperature control is ensured by way of the capillary tube ther-
mostat which is built into the connection box and which has a regulation range between 0 to 180 °C [356 °F] (
-H
)
or 0 to 320 °C [608 °F] (
-H320/C
). Rated values can be set to a maximum 180 °C [356 °F] resp. 320 °C [608 °F]. The
temperature has an excess temperature limiter which permanently switches off the heating if the set temperature
is exceeded by 30 °C. The heating can be switched back on again by pressing the green RESET-button which is
located below the aperture in the thermostats mounting.
The thermostat has a temperature failure alarm which activates a contact if the temperature falls 30 °C below the
set temperature. This status alarm is located on the terminal strip as a potential-free switch over contact.
If temperature control is to be achieved using external electronic equipment, a PT100 (a max. of two can be built
in) or an Fe-CuNi resp. Ni-CrNi thermoelement is provided as a temperature sensor.
M&C
also supplies suitable temperature controller, e.g. Type
70304G
(see data sheet 4.3) which also can be
mounted directly at the gas sample probe (max. ambient tempe45 °C [113 °F]).
11.3
SAMPLE TUBE AND PREFILTER POSSIBILITIES
Depending on the process gas temperature and composition, probe tubes made of different materials are used
with G ¾" connections.
Sample tube
Type
Part No.
Temperature
max. °C
Material
tube/connection part
Length
1)
mm
Length
max.
mm
Tube ø
o/i „d1“
mm
SP2000/PV
20S9070 90
PVDF/
PTFE-GV
1000 1500 25/21
SP32**
20S9280 90
PVDF/
PTFE-GV
800 800 50/44
SP2000/T
20S9083 160
PTFE/
PTFE-GV
500 500 25/15
SP2000/Ti
20S9075 400
Titanium
1000 2500 25/22
SP210/SS
02S9200 600
SS316Ti
1000 2000 12/10
SP2000/SS
20S9065 600
SS316Ti
1000 2500 25/22
SP2000/SS-Vm
20S9067 600
SS316Ti
1000 2500 25/06
SP2000/HC
20S9090 900
Hastelloy
®
C4
1000
2500
25/22
SP2000/KA
20S9080 1300
Kanthal
®
/ SS316Ti
1000
1500
27/20
SP2000/IN
20S9077 1100
Inconel
®
1000
2000
25/22
SP2000/CR-2*
20S9098 1400
Cr
AL
2
O
3
/ Hastelloy
®
C4 900
900
22.5/13
SP2000/CR-20*
20S9099 1400
Cr
AL
2
O
3
/ Hastelloy
®
C4 1200
1200
22.5/13
SP2000/AO
without con-
nector/adapter
20S9385 1800
Aluminiumoxide
2)
/ 1000
1500 24/18
Summary of Contents for SP Series
Page 48: ...48 SP2000 SP2300 SP2400 1 00 03 www mc techgroup com Figure 24 SP2000 H basic version ...
Page 49: ...www mc techgroup com SP2000 SP2300 SP2400 1 00 03 49 Figure 25 SP2000 H with options ...
Page 50: ...50 SP2000 SP2300 SP2400 1 00 03 www mc techgroup com Figure 26 SP2300 H ...
Page 51: ...www mc techgroup com SP2000 SP2300 SP2400 1 00 03 51 Figure 27 SP2300 H filter elements ...
Page 53: ...www mc techgroup com SP2000 SP2300 SP2400 1 00 03 53 Figure 29 Heated sample tubes SP30 H ...
Page 54: ...54 SP2000 SP2300 SP2400 1 00 03 www mc techgroup com ...
Page 56: ...56 SP2000 SP2300 SP2400 1 00 03 www mc techgroup com Figure 31 SP2000 H 3VA MS NC B ...
Page 57: ...www mc techgroup com SP2000 SP2300 SP2400 1 00 03 57 Figure 32 SP2000 H 3VA MS NC C ...
Page 58: ...58 SP2000 SP2300 SP2400 1 00 03 www mc techgroup com Figure 33 SP2000 H 3VA MS NO B ...
Page 59: ...www mc techgroup com SP2000 SP2300 SP2400 1 00 03 59 Figure 34 SP2000 H 3VA MS NO C ...