
www.mc-techgroup.com SP2000,
SP2300, SP2400 | 1.00.03
25
With option
-H320/C
it is possible to feed test gas via the standard connection for a 6mm tube that is equipped
with a blind cap for the measuring operation. It is placed directly below the heat conducting jaws. For the
connection of the calibration gas line, the blind cap must be removed. By means of the union nut that is attached
to the consignment, the calibration gas line can be connected.
Attention
After the test gas feeding being finished, the connection must be shut again with the
blind cap because otherwise this connection would aspirate secondary air and falsify the
measuring result!
Optionally, the test gas feeding or back purging can be effected via a check valve
-H320/R.
The check valve is
mounted in the area of the bottom plate. The connecting line (tube/pipe, 6 mm outside diameter) can be con-
nected directly on the check valve.
After having finished the mountage, put on again the protection cap and fasten it with the quick action locks.
14.6
CONNECTION OPTION PNEUMATIC DRIVE MS1 OR MS3
The drive line for actuating the pneumatic drive is connected via a corresponding tube or pipe fitting with G1/4“
outside thread
(figure 8 and figures 28 to 31). The required drive pressure is 6.5 up to 9 bar abs.
15
ELECTRICAL CONNECTIONS
Warning
When connecting the equipment, please ensure that the supply voltage is identi-
cal with the information provided on the model type plate.
Note
Attention must be paid to the requirements of IEC 364 (DIN VDE 0100) when set-
ting high-power electrical units with nominal voltages of up to 1000 V, together
with the associated standards and stipulations.
In any case we recommend the use of temperature resistant cable !
A main switch and matching fuse must be provided externally!
The main circuit must be equipped with a fuse corresponding to the nominal cur-
rent (over current protection); for electrical details see technical data.
It is recommended to use the low temperature alarm. In case of an alarm the flow
can be stopped and the components downstream the probe are safe for demage.
Summary of Contents for SP Series
Page 48: ...48 SP2000 SP2300 SP2400 1 00 03 www mc techgroup com Figure 24 SP2000 H basic version ...
Page 49: ...www mc techgroup com SP2000 SP2300 SP2400 1 00 03 49 Figure 25 SP2000 H with options ...
Page 50: ...50 SP2000 SP2300 SP2400 1 00 03 www mc techgroup com Figure 26 SP2300 H ...
Page 51: ...www mc techgroup com SP2000 SP2300 SP2400 1 00 03 51 Figure 27 SP2300 H filter elements ...
Page 53: ...www mc techgroup com SP2000 SP2300 SP2400 1 00 03 53 Figure 29 Heated sample tubes SP30 H ...
Page 54: ...54 SP2000 SP2300 SP2400 1 00 03 www mc techgroup com ...
Page 56: ...56 SP2000 SP2300 SP2400 1 00 03 www mc techgroup com Figure 31 SP2000 H 3VA MS NC B ...
Page 57: ...www mc techgroup com SP2000 SP2300 SP2400 1 00 03 57 Figure 32 SP2000 H 3VA MS NC C ...
Page 58: ...58 SP2000 SP2300 SP2400 1 00 03 www mc techgroup com Figure 33 SP2000 H 3VA MS NO B ...
Page 59: ...www mc techgroup com SP2000 SP2300 SP2400 1 00 03 59 Figure 34 SP2000 H 3VA MS NO C ...