www.mc-techgroup.com SP2000,
SP2300, SP2400 | 1.00.03
11
9
APPLICATION
The
M&C
gas sample probes type
SP2000...
,
SP2300-H
and
SP2400-H
are used for continuous extraction of gases
from dust laden, high-temperature and/or humid processes.
10
DESCRIPTION
The sample probes are designed for easy installation, reliable and flexible operation and trouble-free maintenance.
Depending on the problem, different sample tubes or prefilters (see data sheet 2.14 and 2.17) are screwed into the
thread (G 3/4” i) in the mounting flange. This equipment is not included in the scope of delivery of the probe.
The large-surface ceramic deep filter element (fibre glass or glass wadding fillings are also available) is placed in a
housing with small dead volume outside the process area. The probes are designed so that no tools are necessary
for changing the filter element, the sample line needs not to be dismounted and no contamination on the clean
gas side will occur. Cleaning or back purging of the sample tube is possible from outside.
Due to the special execution of the heating element of the
SP2000-H, SP2300-H
and
SP2400-H
(with protection
cover), the complete filter housing including mounting flange is heated, adjustable up to 180 °C [356 °F] (version
H320/C
up to 320 °C [608 °F]), so that a safe operation without shortfall below the dew point in the out-of-process
area is guaranteed. The temperature adjustment of the standard version is made via an integrated capillary sensor
thermostate with exess temperature limiter and alarm function in compact configuration. Calibration gas feeding
and comparison sampling at the probe are possible.
Depending on the gas composition, it may be possible that the standard material of the probe body (stainless steel
1.4404) is not sufficient with respect to ist corrosion resistance. For this case, the
SP2300-H
of PTFE is available or
alternatively for a heating above 180 °C [356 °F] the
SP2400-H
out of titanium.
The following filter elements are available:
Filter elements
Type
Porosity
Material
Filter element
S-2K 150
2 μm
Ceramic
*
Filter element
S-3G 150
3 μm
Glass
Filter element
S-3SS 150
3 μm
Stainless steel SS316
Filter element
S-0,1GF 150
0.1 μm
Glass fiber
Filter element
FW
--------
Spun glass
*
= Standard
Summary of Contents for SP Series
Page 48: ...48 SP2000 SP2300 SP2400 1 00 03 www mc techgroup com Figure 24 SP2000 H basic version ...
Page 49: ...www mc techgroup com SP2000 SP2300 SP2400 1 00 03 49 Figure 25 SP2000 H with options ...
Page 50: ...50 SP2000 SP2300 SP2400 1 00 03 www mc techgroup com Figure 26 SP2300 H ...
Page 51: ...www mc techgroup com SP2000 SP2300 SP2400 1 00 03 51 Figure 27 SP2300 H filter elements ...
Page 53: ...www mc techgroup com SP2000 SP2300 SP2400 1 00 03 53 Figure 29 Heated sample tubes SP30 H ...
Page 54: ...54 SP2000 SP2300 SP2400 1 00 03 www mc techgroup com ...
Page 56: ...56 SP2000 SP2300 SP2400 1 00 03 www mc techgroup com Figure 31 SP2000 H 3VA MS NC B ...
Page 57: ...www mc techgroup com SP2000 SP2300 SP2400 1 00 03 57 Figure 32 SP2000 H 3VA MS NC C ...
Page 58: ...58 SP2000 SP2300 SP2400 1 00 03 www mc techgroup com Figure 33 SP2000 H 3VA MS NO B ...
Page 59: ...www mc techgroup com SP2000 SP2300 SP2400 1 00 03 59 Figure 34 SP2000 H 3VA MS NO C ...