10
SP2000, SP2300, SP2400 | 1.00.03
www.mc-techgroup.com
8
TECHNICAL DATA
Gas sample probe type
SP2000
SP2000-H
SP2300-H
SP2400-H
Part No.
20S1000
20S2000
20S3000
20S3500
Weather protection shield
No
Yes
Yes
Yes
Protection class terminal box
IP 54 EN 60529
Material filter housing
Stainless steel SS316/316Ti*
PTFE
Titanium
Sealing material
FKM* /7aT** = PTFE -H320/C** = Graphite
Probe flange sealing material
Novapress
Insitu probe tube/prefilter Optional
Sample pressure max.
0.4 - 6 bar* abs., /7aT**= 2 bar abs.,
/HP** = 25 bar abs.
2 bar abs.
0.4 - 6 bar abs.
Ambient temperature
20 to +180 °C [68
to 356 °F]
-20 to +60 °C* [-4 to 140 °F];
/PT100, /Fe-CuNi, /Ni-CrNi** = -20 to +80 °C [-4 to 176 °F]
Filter chamber volume
120 cm
3
Filter porosity
S-2K150= ceramic*, 2 μm, /F-0, 1GF150= glass fiber**, 0.1 μm,
/F-3SS150= stainless steel**, 3 μm, /FW= spun glass**
Thermostat, temperature adjust-
ment
0 to 180 °C* [32 to 356 °F] -H320/C**= 0 – 320 °C [0 to 608 °F]
/PT100** /Fe-CuNi** /Ni-CrNi**
Ready for operation
After 40 min -H320/C** = after 60 min
Low temperature alarm contact*
Contact rating: 250 V, 3 A~, 0.25 A= alarm point: ΔT 30°C
Sample gas outlet cnnection
1x 1/4“ NPTi* tube connection** ø 6, 8 or 10 mm
-H320/C**= tube connection 6 or 8 mm*
Backflush/Test gas connection
1/4“ NPTi* /R**, -H320/C**= tube ø 6 mm
Power supply
230 V 50/60 Hz, 800 W /115 V** = 115 V 60 Hz, 800 W (fuse pro-
tection 10 A)
Electrical connections
Terminals max. 4 mm
2
, 2 x cable gland M20 x 1.5
Electrical equipment standard
EN 61010, EN 60519-1
Mounting flange
DN 65 PN 6, form B >DN or ANSI possible** /HP** = DN 50 PN 25
Mounting flange material
SS316Ti
PTFE
Titanium
Weight
7 kg* [15.4 lbs]
15.4 kg* [34 lbs]
15.4 kg*
[34 lbs]
14.5 kg*
[32 lbs]
*
= Standard
** = Optional
Differential pressure and T90-time at different flow rates
P and T90 at flow of
100
200 500 1000
1500
3000
(only/HF)
Nl/h
P with new filter element
S-2K 150/F-0,1GF150
0.007
0.011 0.020 0.058 0.135 0.240/0.225 bar
P with new filter element
F-3SS150
0.006
0.012 0.040 0.110 0.215 0.405
bar
T90-time without sample tube/pre-
filter
6.0
3.5
1.0
< 0.5
< 0.5
< 0.5
s
Summary of Contents for SP Series
Page 48: ...48 SP2000 SP2300 SP2400 1 00 03 www mc techgroup com Figure 24 SP2000 H basic version ...
Page 49: ...www mc techgroup com SP2000 SP2300 SP2400 1 00 03 49 Figure 25 SP2000 H with options ...
Page 50: ...50 SP2000 SP2300 SP2400 1 00 03 www mc techgroup com Figure 26 SP2300 H ...
Page 51: ...www mc techgroup com SP2000 SP2300 SP2400 1 00 03 51 Figure 27 SP2300 H filter elements ...
Page 53: ...www mc techgroup com SP2000 SP2300 SP2400 1 00 03 53 Figure 29 Heated sample tubes SP30 H ...
Page 54: ...54 SP2000 SP2300 SP2400 1 00 03 www mc techgroup com ...
Page 56: ...56 SP2000 SP2300 SP2400 1 00 03 www mc techgroup com Figure 31 SP2000 H 3VA MS NC B ...
Page 57: ...www mc techgroup com SP2000 SP2300 SP2400 1 00 03 57 Figure 32 SP2000 H 3VA MS NC C ...
Page 58: ...58 SP2000 SP2300 SP2400 1 00 03 www mc techgroup com Figure 33 SP2000 H 3VA MS NO B ...
Page 59: ...www mc techgroup com SP2000 SP2300 SP2400 1 00 03 59 Figure 34 SP2000 H 3VA MS NO C ...