Industrial Gaussmeter IGM11 Operating Instructions
Chapter 9 Serial Interfaces
Page 91 / 106
MAGSYS
magne t
systeme
Version 01/2013
© 2013 MAGSYS magnet syste
me GmbH - All Rights Reserved.
9.12.7.12
:DGIO:SET:IN
{x}
:FUNC[?]
Description:
Set or query the function of a digital input.
x is for inputs 1 to 3.
Mode:
Command or query
Parameter:
{
NO|HOLD|NLPK|NULL
}
Parameter
Function
NO
No function assigned to this input
HOLD
Setting this input will stop the measurement
NLPK
Setting this input will reset the peak values (with
rising edge)
NULL
Setting this input will zero compensate the input
(with edge)
Response:
like parameter
*RST
Value:
not relevant
Example:
send
:DGIO:SET:IN2:FUNC NLPK<LF>
send
:DGIO:SET:IN2:FUNC?<LF>
receive NLPK<CR><LF>
9.12.7.13
:DGIO:SET:IN
{x}
:LOG[?]
Description:
Set or query the logic level of a digital input.
x is for inputs 1 to 3.
Mode:
Command or query
Parameter:
{
0|1
}
0 is for normal function (current flow sets the input)
1 is for inverted function
Response:
like parameter
*RST
Value:
not relevant
Example:
send
:DGIO:SET:IN2:LOG 1<LF>
send
:DGIO:SET:IN2:LOG?<LF>
receive 1<CR><LF>